Accuracy of inhomogeneity correction algorithm in intensity-modulated radiotherapy of head-and-neck tumors

Myonggeun Yoon, Doo Hyun Lee, Dongho Shin, Se Byeong Lee, Sung Yong Park, Kwan Ho Cho

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

We examined the degree of calculated-to-measured dose difference for nasopharyngeal target volume in intensity-modulated radiotherapy (IMRT) based on the observed/expected ratio using patient anatomy with humanoid head-and-neck phantom. The plans were designed with a clinical treatment planning system that uses a measurement-based pencil beam dose-calculation algorithm. Two kinds of IMRT plans, which give a direct indication of the error introduced in routine treatment planning, were categorized and evaluated. The experimental results show that when the beams pass through the oral cavity in anthropomorphic head-and-neck phantom, the average dose difference becomes significant, revealing about 10% dose difference to prescribed dose at isocenter. To investigate both the physical reasons of the dose discrepancy and the inhomogeneity effect, we performed the 10 cases of IMRT quality assurance (QA) with plastic and humanoid phantoms. Our result suggests that the transient electronic disequilibrium with the increased lateral electron range may cause the inaccuracy of dose calculation algorithm, and the effectiveness of the inhomogeneity corrections used in IMRT plans should be evaluated to ensure meaningful quality assurance and delivery.

Original languageEnglish
Pages (from-to)44-51
Number of pages8
JournalMedical Dosimetry
Volume32
Issue number1
DOIs
Publication statusPublished - 2007 Mar 1
Externally publishedYes

Fingerprint

Intensity-Modulated Radiotherapy
radiation therapy
inhomogeneity
Neck
tumors
Head
dosage
Neoplasms
assurance
planning
Plastics
Mouth
Anatomy
pencil beams
Electrons
anatomy
Therapeutics
delivery
indication
plastics

Keywords

  • Dosimetry
  • IMRT
  • Inhomogeneity
  • Oral-cavity

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Health Professions(all)
  • Radiation

Cite this

Accuracy of inhomogeneity correction algorithm in intensity-modulated radiotherapy of head-and-neck tumors. / Yoon, Myonggeun; Lee, Doo Hyun; Shin, Dongho; Lee, Se Byeong; Park, Sung Yong; Cho, Kwan Ho.

In: Medical Dosimetry, Vol. 32, No. 1, 01.03.2007, p. 44-51.

Research output: Contribution to journalArticle

Yoon, Myonggeun ; Lee, Doo Hyun ; Shin, Dongho ; Lee, Se Byeong ; Park, Sung Yong ; Cho, Kwan Ho. / Accuracy of inhomogeneity correction algorithm in intensity-modulated radiotherapy of head-and-neck tumors. In: Medical Dosimetry. 2007 ; Vol. 32, No. 1. pp. 44-51.
@article{996471a2ad794cce98c1d0cbdf42f2ec,
title = "Accuracy of inhomogeneity correction algorithm in intensity-modulated radiotherapy of head-and-neck tumors",
abstract = "We examined the degree of calculated-to-measured dose difference for nasopharyngeal target volume in intensity-modulated radiotherapy (IMRT) based on the observed/expected ratio using patient anatomy with humanoid head-and-neck phantom. The plans were designed with a clinical treatment planning system that uses a measurement-based pencil beam dose-calculation algorithm. Two kinds of IMRT plans, which give a direct indication of the error introduced in routine treatment planning, were categorized and evaluated. The experimental results show that when the beams pass through the oral cavity in anthropomorphic head-and-neck phantom, the average dose difference becomes significant, revealing about 10{\%} dose difference to prescribed dose at isocenter. To investigate both the physical reasons of the dose discrepancy and the inhomogeneity effect, we performed the 10 cases of IMRT quality assurance (QA) with plastic and humanoid phantoms. Our result suggests that the transient electronic disequilibrium with the increased lateral electron range may cause the inaccuracy of dose calculation algorithm, and the effectiveness of the inhomogeneity corrections used in IMRT plans should be evaluated to ensure meaningful quality assurance and delivery.",
keywords = "Dosimetry, IMRT, Inhomogeneity, Oral-cavity",
author = "Myonggeun Yoon and Lee, {Doo Hyun} and Dongho Shin and Lee, {Se Byeong} and Park, {Sung Yong} and Cho, {Kwan Ho}",
year = "2007",
month = "3",
day = "1",
doi = "10.1016/j.meddos.2006.11.004",
language = "English",
volume = "32",
pages = "44--51",
journal = "Medical Dosimetry",
issn = "0958-3947",
publisher = "Elsevier Inc.",
number = "1",

}

TY - JOUR

T1 - Accuracy of inhomogeneity correction algorithm in intensity-modulated radiotherapy of head-and-neck tumors

AU - Yoon, Myonggeun

AU - Lee, Doo Hyun

AU - Shin, Dongho

AU - Lee, Se Byeong

AU - Park, Sung Yong

AU - Cho, Kwan Ho

PY - 2007/3/1

Y1 - 2007/3/1

N2 - We examined the degree of calculated-to-measured dose difference for nasopharyngeal target volume in intensity-modulated radiotherapy (IMRT) based on the observed/expected ratio using patient anatomy with humanoid head-and-neck phantom. The plans were designed with a clinical treatment planning system that uses a measurement-based pencil beam dose-calculation algorithm. Two kinds of IMRT plans, which give a direct indication of the error introduced in routine treatment planning, were categorized and evaluated. The experimental results show that when the beams pass through the oral cavity in anthropomorphic head-and-neck phantom, the average dose difference becomes significant, revealing about 10% dose difference to prescribed dose at isocenter. To investigate both the physical reasons of the dose discrepancy and the inhomogeneity effect, we performed the 10 cases of IMRT quality assurance (QA) with plastic and humanoid phantoms. Our result suggests that the transient electronic disequilibrium with the increased lateral electron range may cause the inaccuracy of dose calculation algorithm, and the effectiveness of the inhomogeneity corrections used in IMRT plans should be evaluated to ensure meaningful quality assurance and delivery.

AB - We examined the degree of calculated-to-measured dose difference for nasopharyngeal target volume in intensity-modulated radiotherapy (IMRT) based on the observed/expected ratio using patient anatomy with humanoid head-and-neck phantom. The plans were designed with a clinical treatment planning system that uses a measurement-based pencil beam dose-calculation algorithm. Two kinds of IMRT plans, which give a direct indication of the error introduced in routine treatment planning, were categorized and evaluated. The experimental results show that when the beams pass through the oral cavity in anthropomorphic head-and-neck phantom, the average dose difference becomes significant, revealing about 10% dose difference to prescribed dose at isocenter. To investigate both the physical reasons of the dose discrepancy and the inhomogeneity effect, we performed the 10 cases of IMRT quality assurance (QA) with plastic and humanoid phantoms. Our result suggests that the transient electronic disequilibrium with the increased lateral electron range may cause the inaccuracy of dose calculation algorithm, and the effectiveness of the inhomogeneity corrections used in IMRT plans should be evaluated to ensure meaningful quality assurance and delivery.

KW - Dosimetry

KW - IMRT

KW - Inhomogeneity

KW - Oral-cavity

UR - http://www.scopus.com/inward/record.url?scp=33847084600&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33847084600&partnerID=8YFLogxK

U2 - 10.1016/j.meddos.2006.11.004

DO - 10.1016/j.meddos.2006.11.004

M3 - Article

C2 - 17317535

AN - SCOPUS:33847084600

VL - 32

SP - 44

EP - 51

JO - Medical Dosimetry

JF - Medical Dosimetry

SN - 0958-3947

IS - 1

ER -