Biomechanical assessment of a novel bone lengthening plate system - A cadaveric study

Tae Gon Jung, Seung-Woo Suh, Sung Jae Lee, Bongju Kim, Dong Wook Han, Jae Hyuk Yang

Research output: Contribution to journalArticle

Abstract

Background Although many types of external fixators have been developed for distraction osteogenesis, all have some drawbacks. We recently developed a novel bone lengthening plate to overcome these problems. The purpose of this study is to conduct biomechanical analyses using cadavers to assess the stability of the bone lengthening plate in relation to distraction length and femoral bone mineral density. Methods We used human cadaveric femurs (n = 18) to assess the effects of distraction length and bone mineral density on the biomechanical stability of the bone lengthening plate. After establishing control (n = 6, 0 mm lengthening) and experimental groups (n = 12, 30 mm lengthening), we measured biomechanical stability (structural stiffness, ultimate load, and displacement) under a compressive load. The experimental group was subdivided into a group with normal bone mineral density (n = 6) and a group with osteoporosis (n = 6), and the biomechanical stability of these groups was compared. Finding Structural stiffness differed significantly between the control (417.6 N/mm) and combined experimental groups (185.6 N/mm, p = 0.002). Ultimate load also differed significantly between the control (1327.8 N) and combined experimental (331.4 N, p = 0.002) groups. Bone mineral density was unrelated to structural stiffness (p = 0.204), ultimate load (0.876), or displacement (0.344). In all cases, failure of the bone lengthening plate occurred at the longitudinal connectors, such as the connecting columns between the upper and lower plates, and the lengthening shaft of the bone lengthening plate. Interpretation The biomechanical stability of the bone lengthening plate was affected by the lengthening length but not by bone mineral density. In addition, biomechanical stability during lengthening was most strongly influenced by the longitudinal connectors.

Original languageEnglish
Pages (from-to)232-238
Number of pages7
JournalClinical Biomechanics
Volume28
Issue number2
DOIs
Publication statusPublished - 2013 Feb 1

Fingerprint

Bone Lengthening
Bone Plates
Bone Density
Distraction Osteogenesis
External Fixators
Thigh
Cadaver
Femur
Osteoporosis

Keywords

  • Biomechanical stability
  • Bone lengthening plate
  • Bone mineral density
  • Femur
  • Keywords

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine
  • Biophysics

Cite this

Biomechanical assessment of a novel bone lengthening plate system - A cadaveric study. / Jung, Tae Gon; Suh, Seung-Woo; Lee, Sung Jae; Kim, Bongju; Han, Dong Wook; Yang, Jae Hyuk.

In: Clinical Biomechanics, Vol. 28, No. 2, 01.02.2013, p. 232-238.

Research output: Contribution to journalArticle

Jung, Tae Gon ; Suh, Seung-Woo ; Lee, Sung Jae ; Kim, Bongju ; Han, Dong Wook ; Yang, Jae Hyuk. / Biomechanical assessment of a novel bone lengthening plate system - A cadaveric study. In: Clinical Biomechanics. 2013 ; Vol. 28, No. 2. pp. 232-238.
@article{3d721b49de78463db79a0f936b40bee4,
title = "Biomechanical assessment of a novel bone lengthening plate system - A cadaveric study",
abstract = "Background Although many types of external fixators have been developed for distraction osteogenesis, all have some drawbacks. We recently developed a novel bone lengthening plate to overcome these problems. The purpose of this study is to conduct biomechanical analyses using cadavers to assess the stability of the bone lengthening plate in relation to distraction length and femoral bone mineral density. Methods We used human cadaveric femurs (n = 18) to assess the effects of distraction length and bone mineral density on the biomechanical stability of the bone lengthening plate. After establishing control (n = 6, 0 mm lengthening) and experimental groups (n = 12, 30 mm lengthening), we measured biomechanical stability (structural stiffness, ultimate load, and displacement) under a compressive load. The experimental group was subdivided into a group with normal bone mineral density (n = 6) and a group with osteoporosis (n = 6), and the biomechanical stability of these groups was compared. Finding Structural stiffness differed significantly between the control (417.6 N/mm) and combined experimental groups (185.6 N/mm, p = 0.002). Ultimate load also differed significantly between the control (1327.8 N) and combined experimental (331.4 N, p = 0.002) groups. Bone mineral density was unrelated to structural stiffness (p = 0.204), ultimate load (0.876), or displacement (0.344). In all cases, failure of the bone lengthening plate occurred at the longitudinal connectors, such as the connecting columns between the upper and lower plates, and the lengthening shaft of the bone lengthening plate. Interpretation The biomechanical stability of the bone lengthening plate was affected by the lengthening length but not by bone mineral density. In addition, biomechanical stability during lengthening was most strongly influenced by the longitudinal connectors.",
keywords = "Biomechanical stability, Bone lengthening plate, Bone mineral density, Femur, Keywords",
author = "Jung, {Tae Gon} and Seung-Woo Suh and Lee, {Sung Jae} and Bongju Kim and Han, {Dong Wook} and Yang, {Jae Hyuk}",
year = "2013",
month = "2",
day = "1",
doi = "10.1016/j.clinbiomech.2012.11.011",
language = "English",
volume = "28",
pages = "232--238",
journal = "Clinical Biomechanics",
issn = "0268-0033",
publisher = "Elsevier Limited",
number = "2",

}

TY - JOUR

T1 - Biomechanical assessment of a novel bone lengthening plate system - A cadaveric study

AU - Jung, Tae Gon

AU - Suh, Seung-Woo

AU - Lee, Sung Jae

AU - Kim, Bongju

AU - Han, Dong Wook

AU - Yang, Jae Hyuk

PY - 2013/2/1

Y1 - 2013/2/1

N2 - Background Although many types of external fixators have been developed for distraction osteogenesis, all have some drawbacks. We recently developed a novel bone lengthening plate to overcome these problems. The purpose of this study is to conduct biomechanical analyses using cadavers to assess the stability of the bone lengthening plate in relation to distraction length and femoral bone mineral density. Methods We used human cadaveric femurs (n = 18) to assess the effects of distraction length and bone mineral density on the biomechanical stability of the bone lengthening plate. After establishing control (n = 6, 0 mm lengthening) and experimental groups (n = 12, 30 mm lengthening), we measured biomechanical stability (structural stiffness, ultimate load, and displacement) under a compressive load. The experimental group was subdivided into a group with normal bone mineral density (n = 6) and a group with osteoporosis (n = 6), and the biomechanical stability of these groups was compared. Finding Structural stiffness differed significantly between the control (417.6 N/mm) and combined experimental groups (185.6 N/mm, p = 0.002). Ultimate load also differed significantly between the control (1327.8 N) and combined experimental (331.4 N, p = 0.002) groups. Bone mineral density was unrelated to structural stiffness (p = 0.204), ultimate load (0.876), or displacement (0.344). In all cases, failure of the bone lengthening plate occurred at the longitudinal connectors, such as the connecting columns between the upper and lower plates, and the lengthening shaft of the bone lengthening plate. Interpretation The biomechanical stability of the bone lengthening plate was affected by the lengthening length but not by bone mineral density. In addition, biomechanical stability during lengthening was most strongly influenced by the longitudinal connectors.

AB - Background Although many types of external fixators have been developed for distraction osteogenesis, all have some drawbacks. We recently developed a novel bone lengthening plate to overcome these problems. The purpose of this study is to conduct biomechanical analyses using cadavers to assess the stability of the bone lengthening plate in relation to distraction length and femoral bone mineral density. Methods We used human cadaveric femurs (n = 18) to assess the effects of distraction length and bone mineral density on the biomechanical stability of the bone lengthening plate. After establishing control (n = 6, 0 mm lengthening) and experimental groups (n = 12, 30 mm lengthening), we measured biomechanical stability (structural stiffness, ultimate load, and displacement) under a compressive load. The experimental group was subdivided into a group with normal bone mineral density (n = 6) and a group with osteoporosis (n = 6), and the biomechanical stability of these groups was compared. Finding Structural stiffness differed significantly between the control (417.6 N/mm) and combined experimental groups (185.6 N/mm, p = 0.002). Ultimate load also differed significantly between the control (1327.8 N) and combined experimental (331.4 N, p = 0.002) groups. Bone mineral density was unrelated to structural stiffness (p = 0.204), ultimate load (0.876), or displacement (0.344). In all cases, failure of the bone lengthening plate occurred at the longitudinal connectors, such as the connecting columns between the upper and lower plates, and the lengthening shaft of the bone lengthening plate. Interpretation The biomechanical stability of the bone lengthening plate was affected by the lengthening length but not by bone mineral density. In addition, biomechanical stability during lengthening was most strongly influenced by the longitudinal connectors.

KW - Biomechanical stability

KW - Bone lengthening plate

KW - Bone mineral density

KW - Femur

KW - Keywords

UR - http://www.scopus.com/inward/record.url?scp=84875734090&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84875734090&partnerID=8YFLogxK

U2 - 10.1016/j.clinbiomech.2012.11.011

DO - 10.1016/j.clinbiomech.2012.11.011

M3 - Article

C2 - 23261017

AN - SCOPUS:84875734090

VL - 28

SP - 232

EP - 238

JO - Clinical Biomechanics

JF - Clinical Biomechanics

SN - 0268-0033

IS - 2

ER -