Biomechanical assessment of a novel lengthening plate for distraction osteogenesis

A finite element study

Su Heon Woo, Seung-Woo Suh, Tae Gon Jung, Sung Jae Lee, Dong Wook Han, Jae Hyuk Yang

Research output: Contribution to journalArticle

Abstract

Purpose: The purpose of this study was to describe a novel bone lengthening plate (BLP) and to evaluate the biomechanical performance of the plate before clinical use. Methods: A finite element model was obtained from values of the compression test. It was used to evaluate the performance of the BLP, such as biomechanical stability and structural stiffness, according to materials of the shaft and column used for lengthening (type I: all Titanium-alloys, type II and III: Titanium-alloys and Cobalt-Chrome-alloys, type IV: all Cobalt-Chrome-alloys). Furthermore, Peak von Mises Stress, risk of screw breakage, and micro-motion between the bone and screw were evaluated. Results: Structural stiffness and strain of the finite element model of the femur were similar to cadaveric femurs. Structural stiffness increased from type I (Titanium alloys) to type IV (Cobalt-Chrome-alloys) by at least 128%. First locking screw, which was inserted at the osteotomy site of the proximal femur (proximal to distal direction), showed the highest risk of breakage (75%). Conclusions: This study investigated biomechanical stability of the novel BLP in relation to properties of the supporting structures. It was most biomechanically stable when the column and lengthening shaft were composed of Cobalt-Chrome-alloys (type IV).

Original languageEnglish
Pages (from-to)216-223
Number of pages8
JournalBiomedical Engineering Letters
Volume6
Issue number4
DOIs
Publication statusPublished - 2016 Nov 1

Fingerprint

Cobalt
Bone
Titanium alloys
Stiffness

Keywords

  • Bone Lengthening Plate
  • Femur
  • Finite Element Model

ASJC Scopus subject areas

  • Biomedical Engineering

Cite this

Biomechanical assessment of a novel lengthening plate for distraction osteogenesis : A finite element study. / Woo, Su Heon; Suh, Seung-Woo; Jung, Tae Gon; Lee, Sung Jae; Han, Dong Wook; Yang, Jae Hyuk.

In: Biomedical Engineering Letters, Vol. 6, No. 4, 01.11.2016, p. 216-223.

Research output: Contribution to journalArticle

Woo, Su Heon ; Suh, Seung-Woo ; Jung, Tae Gon ; Lee, Sung Jae ; Han, Dong Wook ; Yang, Jae Hyuk. / Biomechanical assessment of a novel lengthening plate for distraction osteogenesis : A finite element study. In: Biomedical Engineering Letters. 2016 ; Vol. 6, No. 4. pp. 216-223.
@article{792eae69e5e249e4844c2f333568fb56,
title = "Biomechanical assessment of a novel lengthening plate for distraction osteogenesis: A finite element study",
abstract = "Purpose: The purpose of this study was to describe a novel bone lengthening plate (BLP) and to evaluate the biomechanical performance of the plate before clinical use. Methods: A finite element model was obtained from values of the compression test. It was used to evaluate the performance of the BLP, such as biomechanical stability and structural stiffness, according to materials of the shaft and column used for lengthening (type I: all Titanium-alloys, type II and III: Titanium-alloys and Cobalt-Chrome-alloys, type IV: all Cobalt-Chrome-alloys). Furthermore, Peak von Mises Stress, risk of screw breakage, and micro-motion between the bone and screw were evaluated. Results: Structural stiffness and strain of the finite element model of the femur were similar to cadaveric femurs. Structural stiffness increased from type I (Titanium alloys) to type IV (Cobalt-Chrome-alloys) by at least 128{\%}. First locking screw, which was inserted at the osteotomy site of the proximal femur (proximal to distal direction), showed the highest risk of breakage (75{\%}). Conclusions: This study investigated biomechanical stability of the novel BLP in relation to properties of the supporting structures. It was most biomechanically stable when the column and lengthening shaft were composed of Cobalt-Chrome-alloys (type IV).",
keywords = "Bone Lengthening Plate, Femur, Finite Element Model",
author = "Woo, {Su Heon} and Seung-Woo Suh and Jung, {Tae Gon} and Lee, {Sung Jae} and Han, {Dong Wook} and Yang, {Jae Hyuk}",
year = "2016",
month = "11",
day = "1",
doi = "10.1007/s13534-016-0224-4",
language = "English",
volume = "6",
pages = "216--223",
journal = "Biomedical Engineering Letters",
issn = "2093-9868",
publisher = "Springer Verlag",
number = "4",

}

TY - JOUR

T1 - Biomechanical assessment of a novel lengthening plate for distraction osteogenesis

T2 - A finite element study

AU - Woo, Su Heon

AU - Suh, Seung-Woo

AU - Jung, Tae Gon

AU - Lee, Sung Jae

AU - Han, Dong Wook

AU - Yang, Jae Hyuk

PY - 2016/11/1

Y1 - 2016/11/1

N2 - Purpose: The purpose of this study was to describe a novel bone lengthening plate (BLP) and to evaluate the biomechanical performance of the plate before clinical use. Methods: A finite element model was obtained from values of the compression test. It was used to evaluate the performance of the BLP, such as biomechanical stability and structural stiffness, according to materials of the shaft and column used for lengthening (type I: all Titanium-alloys, type II and III: Titanium-alloys and Cobalt-Chrome-alloys, type IV: all Cobalt-Chrome-alloys). Furthermore, Peak von Mises Stress, risk of screw breakage, and micro-motion between the bone and screw were evaluated. Results: Structural stiffness and strain of the finite element model of the femur were similar to cadaveric femurs. Structural stiffness increased from type I (Titanium alloys) to type IV (Cobalt-Chrome-alloys) by at least 128%. First locking screw, which was inserted at the osteotomy site of the proximal femur (proximal to distal direction), showed the highest risk of breakage (75%). Conclusions: This study investigated biomechanical stability of the novel BLP in relation to properties of the supporting structures. It was most biomechanically stable when the column and lengthening shaft were composed of Cobalt-Chrome-alloys (type IV).

AB - Purpose: The purpose of this study was to describe a novel bone lengthening plate (BLP) and to evaluate the biomechanical performance of the plate before clinical use. Methods: A finite element model was obtained from values of the compression test. It was used to evaluate the performance of the BLP, such as biomechanical stability and structural stiffness, according to materials of the shaft and column used for lengthening (type I: all Titanium-alloys, type II and III: Titanium-alloys and Cobalt-Chrome-alloys, type IV: all Cobalt-Chrome-alloys). Furthermore, Peak von Mises Stress, risk of screw breakage, and micro-motion between the bone and screw were evaluated. Results: Structural stiffness and strain of the finite element model of the femur were similar to cadaveric femurs. Structural stiffness increased from type I (Titanium alloys) to type IV (Cobalt-Chrome-alloys) by at least 128%. First locking screw, which was inserted at the osteotomy site of the proximal femur (proximal to distal direction), showed the highest risk of breakage (75%). Conclusions: This study investigated biomechanical stability of the novel BLP in relation to properties of the supporting structures. It was most biomechanically stable when the column and lengthening shaft were composed of Cobalt-Chrome-alloys (type IV).

KW - Bone Lengthening Plate

KW - Femur

KW - Finite Element Model

UR - http://www.scopus.com/inward/record.url?scp=85004065486&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85004065486&partnerID=8YFLogxK

U2 - 10.1007/s13534-016-0224-4

DO - 10.1007/s13534-016-0224-4

M3 - Article

VL - 6

SP - 216

EP - 223

JO - Biomedical Engineering Letters

JF - Biomedical Engineering Letters

SN - 2093-9868

IS - 4

ER -