Calibration of Ir-192 HDR source based on standards of absorbed dose to water

Hyun Do Huh, Woo Chul Kim, John J K Loh, Suk Lee, Dong Oh Shin, Sang Hoon Lee, Sangwook Lim, Sooil Kwon, Jinho Choi, Seong Hoon Kim

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

Brachytherapy sources are usually calibrated in terms of the measured air-kerma strength SK in free space at 1 m distance. The trend of the protocols for the external beam radiotherapy has been changing from an air_kerma base to standards of absorbed dose to water. In this work we tried to specify the source strength of Ir-192 source in terms of the absorbed_dose_to_water instead of an air kerma strength using standards of absorbed dose to water. For this work the MPBP(the Multi Purpose Brachytherapy Phantom) was home-made, which was designed to put both sources and chambers into water at the exact point we want. The reference point was chosen at 5 cm distance along the transverse axis of the source. The computer system of Plato RTP, BPS v13.2, Nucletron, Holland, was used to deliver the dose we want to the reference point. The absorbed_ dose_to_water calibration factor NDW of ionization chambers for brachytherapy sources used must be determined in order to apply standards of absorbed dose to water. The PTW TN 30013 chamber was used and the calibration factor NDW of the chamber was determined using the Monte Carlo Code of EGSnrcCPP. The factor calculated was then used to determine the absorbed dose to water, which calculated an air kerma strength for an Ir-192 source at the reference point. The dose calculated was compared with the dose delivered by Plato RTP. The dose delivered to the reference point by the computer system used was 100 cGy. The calibration factor NDW calculated was 5.28 cGy /nC and was used to calculate the absorbed doses at some distances, which show 2.16% to 0.84% difference from the dose delivered. Air kerma strength calculated showed the difference from -2.7% to +1.7% relative to the one of the source provided by a vendor. This shows the feasibility of using standards of absorbed dose to water to calibrate the brachytherapy source.

Original languageEnglish
Title of host publicationIFMBE Proceedings
PublisherSpringer Verlag
Pages1877-1879
Number of pages3
Volume14
Edition1
Publication statusPublished - 2007
Event10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of
Duration: 2006 Aug 272006 Sep 1

Other

Other10th World Congress on Medical Physics and Biomedical Engineering, WC 2006
CountryKorea, Republic of
CitySeoul
Period06/8/2706/9/1

Fingerprint

Calibration
Water
Air
Computer systems
Ionization chambers
Radiotherapy
Dosimetry

Keywords

  • Absorbed dose to water calibration factor
  • Air kerma strength
  • Brachytherapy source

ASJC Scopus subject areas

  • Biomedical Engineering
  • Bioengineering

Cite this

Huh, H. D., Kim, W. C., Loh, J. J. K., Lee, S., Shin, D. O., Lee, S. H., ... Kim, S. H. (2007). Calibration of Ir-192 HDR source based on standards of absorbed dose to water. In IFMBE Proceedings (1 ed., Vol. 14, pp. 1877-1879). Springer Verlag.

Calibration of Ir-192 HDR source based on standards of absorbed dose to water. / Huh, Hyun Do; Kim, Woo Chul; Loh, John J K; Lee, Suk; Shin, Dong Oh; Lee, Sang Hoon; Lim, Sangwook; Kwon, Sooil; Choi, Jinho; Kim, Seong Hoon.

IFMBE Proceedings. Vol. 14 1. ed. Springer Verlag, 2007. p. 1877-1879.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Huh, HD, Kim, WC, Loh, JJK, Lee, S, Shin, DO, Lee, SH, Lim, S, Kwon, S, Choi, J & Kim, SH 2007, Calibration of Ir-192 HDR source based on standards of absorbed dose to water. in IFMBE Proceedings. 1 edn, vol. 14, Springer Verlag, pp. 1877-1879, 10th World Congress on Medical Physics and Biomedical Engineering, WC 2006, Seoul, Korea, Republic of, 06/8/27.
Huh HD, Kim WC, Loh JJK, Lee S, Shin DO, Lee SH et al. Calibration of Ir-192 HDR source based on standards of absorbed dose to water. In IFMBE Proceedings. 1 ed. Vol. 14. Springer Verlag. 2007. p. 1877-1879
Huh, Hyun Do ; Kim, Woo Chul ; Loh, John J K ; Lee, Suk ; Shin, Dong Oh ; Lee, Sang Hoon ; Lim, Sangwook ; Kwon, Sooil ; Choi, Jinho ; Kim, Seong Hoon. / Calibration of Ir-192 HDR source based on standards of absorbed dose to water. IFMBE Proceedings. Vol. 14 1. ed. Springer Verlag, 2007. pp. 1877-1879
@inproceedings{777778b559264be0bb40e0765bc09494,
title = "Calibration of Ir-192 HDR source based on standards of absorbed dose to water",
abstract = "Brachytherapy sources are usually calibrated in terms of the measured air-kerma strength SK in free space at 1 m distance. The trend of the protocols for the external beam radiotherapy has been changing from an air_kerma base to standards of absorbed dose to water. In this work we tried to specify the source strength of Ir-192 source in terms of the absorbed_dose_to_water instead of an air kerma strength using standards of absorbed dose to water. For this work the MPBP(the Multi Purpose Brachytherapy Phantom) was home-made, which was designed to put both sources and chambers into water at the exact point we want. The reference point was chosen at 5 cm distance along the transverse axis of the source. The computer system of Plato RTP, BPS v13.2, Nucletron, Holland, was used to deliver the dose we want to the reference point. The absorbed_ dose_to_water calibration factor NDW of ionization chambers for brachytherapy sources used must be determined in order to apply standards of absorbed dose to water. The PTW TN 30013 chamber was used and the calibration factor NDW of the chamber was determined using the Monte Carlo Code of EGSnrcCPP. The factor calculated was then used to determine the absorbed dose to water, which calculated an air kerma strength for an Ir-192 source at the reference point. The dose calculated was compared with the dose delivered by Plato RTP. The dose delivered to the reference point by the computer system used was 100 cGy. The calibration factor NDW calculated was 5.28 cGy /nC and was used to calculate the absorbed doses at some distances, which show 2.16{\%} to 0.84{\%} difference from the dose delivered. Air kerma strength calculated showed the difference from -2.7{\%} to +1.7{\%} relative to the one of the source provided by a vendor. This shows the feasibility of using standards of absorbed dose to water to calibrate the brachytherapy source.",
keywords = "Absorbed dose to water calibration factor, Air kerma strength, Brachytherapy source",
author = "Huh, {Hyun Do} and Kim, {Woo Chul} and Loh, {John J K} and Suk Lee and Shin, {Dong Oh} and Lee, {Sang Hoon} and Sangwook Lim and Sooil Kwon and Jinho Choi and Kim, {Seong Hoon}",
year = "2007",
language = "English",
volume = "14",
pages = "1877--1879",
booktitle = "IFMBE Proceedings",
publisher = "Springer Verlag",
edition = "1",

}

TY - GEN

T1 - Calibration of Ir-192 HDR source based on standards of absorbed dose to water

AU - Huh, Hyun Do

AU - Kim, Woo Chul

AU - Loh, John J K

AU - Lee, Suk

AU - Shin, Dong Oh

AU - Lee, Sang Hoon

AU - Lim, Sangwook

AU - Kwon, Sooil

AU - Choi, Jinho

AU - Kim, Seong Hoon

PY - 2007

Y1 - 2007

N2 - Brachytherapy sources are usually calibrated in terms of the measured air-kerma strength SK in free space at 1 m distance. The trend of the protocols for the external beam radiotherapy has been changing from an air_kerma base to standards of absorbed dose to water. In this work we tried to specify the source strength of Ir-192 source in terms of the absorbed_dose_to_water instead of an air kerma strength using standards of absorbed dose to water. For this work the MPBP(the Multi Purpose Brachytherapy Phantom) was home-made, which was designed to put both sources and chambers into water at the exact point we want. The reference point was chosen at 5 cm distance along the transverse axis of the source. The computer system of Plato RTP, BPS v13.2, Nucletron, Holland, was used to deliver the dose we want to the reference point. The absorbed_ dose_to_water calibration factor NDW of ionization chambers for brachytherapy sources used must be determined in order to apply standards of absorbed dose to water. The PTW TN 30013 chamber was used and the calibration factor NDW of the chamber was determined using the Monte Carlo Code of EGSnrcCPP. The factor calculated was then used to determine the absorbed dose to water, which calculated an air kerma strength for an Ir-192 source at the reference point. The dose calculated was compared with the dose delivered by Plato RTP. The dose delivered to the reference point by the computer system used was 100 cGy. The calibration factor NDW calculated was 5.28 cGy /nC and was used to calculate the absorbed doses at some distances, which show 2.16% to 0.84% difference from the dose delivered. Air kerma strength calculated showed the difference from -2.7% to +1.7% relative to the one of the source provided by a vendor. This shows the feasibility of using standards of absorbed dose to water to calibrate the brachytherapy source.

AB - Brachytherapy sources are usually calibrated in terms of the measured air-kerma strength SK in free space at 1 m distance. The trend of the protocols for the external beam radiotherapy has been changing from an air_kerma base to standards of absorbed dose to water. In this work we tried to specify the source strength of Ir-192 source in terms of the absorbed_dose_to_water instead of an air kerma strength using standards of absorbed dose to water. For this work the MPBP(the Multi Purpose Brachytherapy Phantom) was home-made, which was designed to put both sources and chambers into water at the exact point we want. The reference point was chosen at 5 cm distance along the transverse axis of the source. The computer system of Plato RTP, BPS v13.2, Nucletron, Holland, was used to deliver the dose we want to the reference point. The absorbed_ dose_to_water calibration factor NDW of ionization chambers for brachytherapy sources used must be determined in order to apply standards of absorbed dose to water. The PTW TN 30013 chamber was used and the calibration factor NDW of the chamber was determined using the Monte Carlo Code of EGSnrcCPP. The factor calculated was then used to determine the absorbed dose to water, which calculated an air kerma strength for an Ir-192 source at the reference point. The dose calculated was compared with the dose delivered by Plato RTP. The dose delivered to the reference point by the computer system used was 100 cGy. The calibration factor NDW calculated was 5.28 cGy /nC and was used to calculate the absorbed doses at some distances, which show 2.16% to 0.84% difference from the dose delivered. Air kerma strength calculated showed the difference from -2.7% to +1.7% relative to the one of the source provided by a vendor. This shows the feasibility of using standards of absorbed dose to water to calibrate the brachytherapy source.

KW - Absorbed dose to water calibration factor

KW - Air kerma strength

KW - Brachytherapy source

UR - http://www.scopus.com/inward/record.url?scp=84958246604&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84958246604&partnerID=8YFLogxK

M3 - Conference contribution

VL - 14

SP - 1877

EP - 1879

BT - IFMBE Proceedings

PB - Springer Verlag

ER -