Clinical application of glass dosimeter for in vivo dose measurements of total body irradiation treatment technique

Jeong Eun Rah, Ui Jung Hwang, Hojin Jeong, Sang Yeob Lee, Doo Hyun Lee, Dong Ho Shin, Myonggeun Yoon, Se Byeong Lee, Rena Lee, Sung Yong Park

Research output: Contribution to journalArticle

21 Citations (Scopus)

Abstract

The commercially available glass dosimeter (model GD-301) was investigated for its dosimetric characteristics, in order to evaluate its use for in vivo dosimetry. We specifically assessed overall precision of dosimetric dose data in patients who received treatment with the total body irradiation (TBI). Uniformity obtained in this study was within 1.2% (1 SD). The dose-response was linear in the range of 0.5-10 Gy with R of 0.999. Dose rate, SSD, field size, angular and energy dependence were found to be within 3.0%. In vivo skin dosimetry for TBI was performed for 3 patients. For all patients, the glass dosimeter was exposed and measured dose recorded for one fraction in addition to conventional used TLD and MOSFET. Overall uncertainty of the glass dosimeter for in vivo dose measurement was estimated at 2.4% (68.3% confidence level). The measured doses of the glass dosimeter were well within ±5.0% of the prescription dose at all sites expect mediastinum of one patient, for which it is within ±5.7%. Agreement of measured doses between glass dosimeter and TLD, MOSFET was within ±6.3% and ±6.6%, respectively. Results show that the glass dosimeter can be used as an accurate and reproducible dosimeter for TBI treatment skin dose measurements. The glass dosimeter is a practical alternative to TLD or MOSFET as an in vivo dosimeter.

Original languageEnglish
Pages (from-to)40-45
Number of pages6
JournalRadiation Measurements
Volume46
Issue number1
DOIs
Publication statusPublished - 2011 Jan 1
Externally publishedYes

Fingerprint

Dosimeters
Dosimetry
dosimeters
Irradiation
Glass
dosage
irradiation
glass
field effect transistors
Skin
mediastinum
confidence

Keywords

  • Glass dosimeter
  • In vivo dosimetry
  • MOSFET
  • TBI (total body irradiation)
  • TLD

ASJC Scopus subject areas

  • Radiation
  • Instrumentation

Cite this

Clinical application of glass dosimeter for in vivo dose measurements of total body irradiation treatment technique. / Rah, Jeong Eun; Hwang, Ui Jung; Jeong, Hojin; Lee, Sang Yeob; Lee, Doo Hyun; Shin, Dong Ho; Yoon, Myonggeun; Lee, Se Byeong; Lee, Rena; Park, Sung Yong.

In: Radiation Measurements, Vol. 46, No. 1, 01.01.2011, p. 40-45.

Research output: Contribution to journalArticle

Rah, Jeong Eun ; Hwang, Ui Jung ; Jeong, Hojin ; Lee, Sang Yeob ; Lee, Doo Hyun ; Shin, Dong Ho ; Yoon, Myonggeun ; Lee, Se Byeong ; Lee, Rena ; Park, Sung Yong. / Clinical application of glass dosimeter for in vivo dose measurements of total body irradiation treatment technique. In: Radiation Measurements. 2011 ; Vol. 46, No. 1. pp. 40-45.
@article{8dbfa702ae23402595b6ab4d262c6899,
title = "Clinical application of glass dosimeter for in vivo dose measurements of total body irradiation treatment technique",
abstract = "The commercially available glass dosimeter (model GD-301) was investigated for its dosimetric characteristics, in order to evaluate its use for in vivo dosimetry. We specifically assessed overall precision of dosimetric dose data in patients who received treatment with the total body irradiation (TBI). Uniformity obtained in this study was within 1.2{\%} (1 SD). The dose-response was linear in the range of 0.5-10 Gy with R of 0.999. Dose rate, SSD, field size, angular and energy dependence were found to be within 3.0{\%}. In vivo skin dosimetry for TBI was performed for 3 patients. For all patients, the glass dosimeter was exposed and measured dose recorded for one fraction in addition to conventional used TLD and MOSFET. Overall uncertainty of the glass dosimeter for in vivo dose measurement was estimated at 2.4{\%} (68.3{\%} confidence level). The measured doses of the glass dosimeter were well within ±5.0{\%} of the prescription dose at all sites expect mediastinum of one patient, for which it is within ±5.7{\%}. Agreement of measured doses between glass dosimeter and TLD, MOSFET was within ±6.3{\%} and ±6.6{\%}, respectively. Results show that the glass dosimeter can be used as an accurate and reproducible dosimeter for TBI treatment skin dose measurements. The glass dosimeter is a practical alternative to TLD or MOSFET as an in vivo dosimeter.",
keywords = "Glass dosimeter, In vivo dosimetry, MOSFET, TBI (total body irradiation), TLD",
author = "Rah, {Jeong Eun} and Hwang, {Ui Jung} and Hojin Jeong and Lee, {Sang Yeob} and Lee, {Doo Hyun} and Shin, {Dong Ho} and Myonggeun Yoon and Lee, {Se Byeong} and Rena Lee and Park, {Sung Yong}",
year = "2011",
month = "1",
day = "1",
doi = "10.1016/j.radmeas.2010.08.013",
language = "English",
volume = "46",
pages = "40--45",
journal = "Radiation Measurements",
issn = "1350-4487",
publisher = "Elsevier Limited",
number = "1",

}

TY - JOUR

T1 - Clinical application of glass dosimeter for in vivo dose measurements of total body irradiation treatment technique

AU - Rah, Jeong Eun

AU - Hwang, Ui Jung

AU - Jeong, Hojin

AU - Lee, Sang Yeob

AU - Lee, Doo Hyun

AU - Shin, Dong Ho

AU - Yoon, Myonggeun

AU - Lee, Se Byeong

AU - Lee, Rena

AU - Park, Sung Yong

PY - 2011/1/1

Y1 - 2011/1/1

N2 - The commercially available glass dosimeter (model GD-301) was investigated for its dosimetric characteristics, in order to evaluate its use for in vivo dosimetry. We specifically assessed overall precision of dosimetric dose data in patients who received treatment with the total body irradiation (TBI). Uniformity obtained in this study was within 1.2% (1 SD). The dose-response was linear in the range of 0.5-10 Gy with R of 0.999. Dose rate, SSD, field size, angular and energy dependence were found to be within 3.0%. In vivo skin dosimetry for TBI was performed for 3 patients. For all patients, the glass dosimeter was exposed and measured dose recorded for one fraction in addition to conventional used TLD and MOSFET. Overall uncertainty of the glass dosimeter for in vivo dose measurement was estimated at 2.4% (68.3% confidence level). The measured doses of the glass dosimeter were well within ±5.0% of the prescription dose at all sites expect mediastinum of one patient, for which it is within ±5.7%. Agreement of measured doses between glass dosimeter and TLD, MOSFET was within ±6.3% and ±6.6%, respectively. Results show that the glass dosimeter can be used as an accurate and reproducible dosimeter for TBI treatment skin dose measurements. The glass dosimeter is a practical alternative to TLD or MOSFET as an in vivo dosimeter.

AB - The commercially available glass dosimeter (model GD-301) was investigated for its dosimetric characteristics, in order to evaluate its use for in vivo dosimetry. We specifically assessed overall precision of dosimetric dose data in patients who received treatment with the total body irradiation (TBI). Uniformity obtained in this study was within 1.2% (1 SD). The dose-response was linear in the range of 0.5-10 Gy with R of 0.999. Dose rate, SSD, field size, angular and energy dependence were found to be within 3.0%. In vivo skin dosimetry for TBI was performed for 3 patients. For all patients, the glass dosimeter was exposed and measured dose recorded for one fraction in addition to conventional used TLD and MOSFET. Overall uncertainty of the glass dosimeter for in vivo dose measurement was estimated at 2.4% (68.3% confidence level). The measured doses of the glass dosimeter were well within ±5.0% of the prescription dose at all sites expect mediastinum of one patient, for which it is within ±5.7%. Agreement of measured doses between glass dosimeter and TLD, MOSFET was within ±6.3% and ±6.6%, respectively. Results show that the glass dosimeter can be used as an accurate and reproducible dosimeter for TBI treatment skin dose measurements. The glass dosimeter is a practical alternative to TLD or MOSFET as an in vivo dosimeter.

KW - Glass dosimeter

KW - In vivo dosimetry

KW - MOSFET

KW - TBI (total body irradiation)

KW - TLD

UR - http://www.scopus.com/inward/record.url?scp=78650844683&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=78650844683&partnerID=8YFLogxK

U2 - 10.1016/j.radmeas.2010.08.013

DO - 10.1016/j.radmeas.2010.08.013

M3 - Article

AN - SCOPUS:78650844683

VL - 46

SP - 40

EP - 45

JO - Radiation Measurements

JF - Radiation Measurements

SN - 1350-4487

IS - 1