Design and QoS of a wireless system for real-time remote electrocardiography

Kyungtae Kang, Junhee Ryu, Junbeom Hur, Lui Sha

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

Quality of service (QoS) and, in particular, reliability and a bounded low latency are essential attributes of safety-critical wireless systems for medical applications. However, wireless links are typically prone to bursts of errors, with characteristics which vary over time.We propose a wireless system suitable for real-time remote patient monitoring in which the necessary reliability and guaranteed latency are both achieved by an efficient error control scheme. We have paired an example remote electrocardiography application to this wireless system. We also developed a tool chain that uses a formal description of the proposed wireless medical system architecture in the architecture analysis and design language to assess various combinations of system parameters: we can determine the QoS in terms of packet-delivery ratio and the service latency, and also the size of jitter buffer required for seamless ECG monitoring. A realistic assessment, based on data from the MIT-BIT arrhythmia database, shows that the proposed wireless system can achieve an appropriate level of QoS for real-time ECG monitoring if link-level error control is correctly implemented. Additionally, we present guidelines for the design of energy-efficient link-level error control, derived from energy data, obtained from simulations.

Original languageEnglish
Pages (from-to)745-755
Number of pages11
JournalIEEE Journal of Biomedical and Health Informatics
Volume17
Issue number3
DOIs
Publication statusPublished - 2013 Oct 14
Externally publishedYes

Fingerprint

Computer Systems
Electrocardiography
Quality of service
Physiologic Monitoring
Cardiac Arrhythmias
Buffers
Remote patient monitoring
Language
Databases
Guidelines
Safety
Monitoring
Medical applications
Jitter
Telecommunication links

Keywords

  • Latency
  • Real-time ECG monitoring
  • Reliability
  • Wireless healthcare

ASJC Scopus subject areas

  • Biotechnology
  • Computer Science Applications
  • Electrical and Electronic Engineering
  • Health Information Management

Cite this

Design and QoS of a wireless system for real-time remote electrocardiography. / Kang, Kyungtae; Ryu, Junhee; Hur, Junbeom; Sha, Lui.

In: IEEE Journal of Biomedical and Health Informatics, Vol. 17, No. 3, 14.10.2013, p. 745-755.

Research output: Contribution to journalArticle

@article{adacba025f6a4a6f8d9c137e8bf8cff3,
title = "Design and QoS of a wireless system for real-time remote electrocardiography",
abstract = "Quality of service (QoS) and, in particular, reliability and a bounded low latency are essential attributes of safety-critical wireless systems for medical applications. However, wireless links are typically prone to bursts of errors, with characteristics which vary over time.We propose a wireless system suitable for real-time remote patient monitoring in which the necessary reliability and guaranteed latency are both achieved by an efficient error control scheme. We have paired an example remote electrocardiography application to this wireless system. We also developed a tool chain that uses a formal description of the proposed wireless medical system architecture in the architecture analysis and design language to assess various combinations of system parameters: we can determine the QoS in terms of packet-delivery ratio and the service latency, and also the size of jitter buffer required for seamless ECG monitoring. A realistic assessment, based on data from the MIT-BIT arrhythmia database, shows that the proposed wireless system can achieve an appropriate level of QoS for real-time ECG monitoring if link-level error control is correctly implemented. Additionally, we present guidelines for the design of energy-efficient link-level error control, derived from energy data, obtained from simulations.",
keywords = "Latency, Real-time ECG monitoring, Reliability, Wireless healthcare",
author = "Kyungtae Kang and Junhee Ryu and Junbeom Hur and Lui Sha",
year = "2013",
month = "10",
day = "14",
doi = "10.1109/JBHI.2013.2237782",
language = "English",
volume = "17",
pages = "745--755",
journal = "IEEE Journal of Biomedical and Health Informatics",
issn = "2168-2194",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "3",

}

TY - JOUR

T1 - Design and QoS of a wireless system for real-time remote electrocardiography

AU - Kang, Kyungtae

AU - Ryu, Junhee

AU - Hur, Junbeom

AU - Sha, Lui

PY - 2013/10/14

Y1 - 2013/10/14

N2 - Quality of service (QoS) and, in particular, reliability and a bounded low latency are essential attributes of safety-critical wireless systems for medical applications. However, wireless links are typically prone to bursts of errors, with characteristics which vary over time.We propose a wireless system suitable for real-time remote patient monitoring in which the necessary reliability and guaranteed latency are both achieved by an efficient error control scheme. We have paired an example remote electrocardiography application to this wireless system. We also developed a tool chain that uses a formal description of the proposed wireless medical system architecture in the architecture analysis and design language to assess various combinations of system parameters: we can determine the QoS in terms of packet-delivery ratio and the service latency, and also the size of jitter buffer required for seamless ECG monitoring. A realistic assessment, based on data from the MIT-BIT arrhythmia database, shows that the proposed wireless system can achieve an appropriate level of QoS for real-time ECG monitoring if link-level error control is correctly implemented. Additionally, we present guidelines for the design of energy-efficient link-level error control, derived from energy data, obtained from simulations.

AB - Quality of service (QoS) and, in particular, reliability and a bounded low latency are essential attributes of safety-critical wireless systems for medical applications. However, wireless links are typically prone to bursts of errors, with characteristics which vary over time.We propose a wireless system suitable for real-time remote patient monitoring in which the necessary reliability and guaranteed latency are both achieved by an efficient error control scheme. We have paired an example remote electrocardiography application to this wireless system. We also developed a tool chain that uses a formal description of the proposed wireless medical system architecture in the architecture analysis and design language to assess various combinations of system parameters: we can determine the QoS in terms of packet-delivery ratio and the service latency, and also the size of jitter buffer required for seamless ECG monitoring. A realistic assessment, based on data from the MIT-BIT arrhythmia database, shows that the proposed wireless system can achieve an appropriate level of QoS for real-time ECG monitoring if link-level error control is correctly implemented. Additionally, we present guidelines for the design of energy-efficient link-level error control, derived from energy data, obtained from simulations.

KW - Latency

KW - Real-time ECG monitoring

KW - Reliability

KW - Wireless healthcare

UR - http://www.scopus.com/inward/record.url?scp=84885095024&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84885095024&partnerID=8YFLogxK

U2 - 10.1109/JBHI.2013.2237782

DO - 10.1109/JBHI.2013.2237782

M3 - Article

VL - 17

SP - 745

EP - 755

JO - IEEE Journal of Biomedical and Health Informatics

JF - IEEE Journal of Biomedical and Health Informatics

SN - 2168-2194

IS - 3

ER -