Detector characteristics of the Micro Crystal Element Scanner (MiCES)

Kisung Lee, Robert S. Miyaoka, Marie L. Janes, Tom K. Lewellen

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

This work reports on the performance characteristics of detector modules for the micro crystal element scanner (MiCES). MiCES is a small animal PET scanner that is designed for imaging mice. It will consist of four rings of 18 detector modules. The crystal array for each detector is a 22 × 22 grid of 0.8 × 0.8 × 10 mm mixed lutetium silicate (MLS) crystals set at 0.9 mm centers and framed for rigidity to compose a 19.94 mm square. A technique to mount multi-element crystal arrays to position sensitive photomultiplier tubes (PSPMTs) is described. The crystal arrays are slightly larger than the dimensions of the anode strips of the PSPMT. Therefore the detector module assembly requires precise alignment of the crystals with the PMT optics for optimum decoding. The crystal arrays are glued to the PSPMT using Histomount (National Diagnostics) glue. It has excellent transmittance at MLS' peak emission wavelength. In addition, tapping down the crystals while the glue was curing was found to significantly improve the amount of light collected. We introduce a semi-automatic boundary segmentation technique to generate look-up-tables used for decoding the crystal of interaction from the collected PSPMT signals. The algorithm is based on valley detection followed by line segment tracking. The software tool provides a more convenient and precise way to create the crystal map for each detector module. Performance characteristics of each of the first 18 detector modules that were assembled and processed with the described techniques are presented. The overall results are an average energy resolution of 23.8% and an average peak-to-valley ratio of 7.6. These averages, however, include modules that had the crystal arrays removed and re-glued multiple times as part of our learning curve of how to optimize the gluing and positioning procedures.

Original languageEnglish
Pages (from-to)1428-1433
Number of pages6
JournalIEEE Transactions on Nuclear Science
Volume52
Issue number5 I
DOIs
Publication statusPublished - 2005 Oct 1
Externally publishedYes

Fingerprint

scanners
Detectors
Crystals
detectors
crystals
modules
Photomultipliers
photomultiplier tubes
Lutetium
lutetium
glues
Glues
decoding
Silicates
valleys
Decoding
silicates
learning curves
Gluing
software development tools

Keywords

  • Detector
  • High resolution
  • Positron emission tomography (PET)
  • Small animal

ASJC Scopus subject areas

  • Electrical and Electronic Engineering
  • Nuclear Energy and Engineering

Cite this

Detector characteristics of the Micro Crystal Element Scanner (MiCES). / Lee, Kisung; Miyaoka, Robert S.; Janes, Marie L.; Lewellen, Tom K.

In: IEEE Transactions on Nuclear Science, Vol. 52, No. 5 I, 01.10.2005, p. 1428-1433.

Research output: Contribution to journalArticle

Lee, Kisung ; Miyaoka, Robert S. ; Janes, Marie L. ; Lewellen, Tom K. / Detector characteristics of the Micro Crystal Element Scanner (MiCES). In: IEEE Transactions on Nuclear Science. 2005 ; Vol. 52, No. 5 I. pp. 1428-1433.
@article{202ecc6893284486bcd313c8b05079fe,
title = "Detector characteristics of the Micro Crystal Element Scanner (MiCES)",
abstract = "This work reports on the performance characteristics of detector modules for the micro crystal element scanner (MiCES). MiCES is a small animal PET scanner that is designed for imaging mice. It will consist of four rings of 18 detector modules. The crystal array for each detector is a 22 × 22 grid of 0.8 × 0.8 × 10 mm mixed lutetium silicate (MLS) crystals set at 0.9 mm centers and framed for rigidity to compose a 19.94 mm square. A technique to mount multi-element crystal arrays to position sensitive photomultiplier tubes (PSPMTs) is described. The crystal arrays are slightly larger than the dimensions of the anode strips of the PSPMT. Therefore the detector module assembly requires precise alignment of the crystals with the PMT optics for optimum decoding. The crystal arrays are glued to the PSPMT using Histomount (National Diagnostics) glue. It has excellent transmittance at MLS' peak emission wavelength. In addition, tapping down the crystals while the glue was curing was found to significantly improve the amount of light collected. We introduce a semi-automatic boundary segmentation technique to generate look-up-tables used for decoding the crystal of interaction from the collected PSPMT signals. The algorithm is based on valley detection followed by line segment tracking. The software tool provides a more convenient and precise way to create the crystal map for each detector module. Performance characteristics of each of the first 18 detector modules that were assembled and processed with the described techniques are presented. The overall results are an average energy resolution of 23.8{\%} and an average peak-to-valley ratio of 7.6. These averages, however, include modules that had the crystal arrays removed and re-glued multiple times as part of our learning curve of how to optimize the gluing and positioning procedures.",
keywords = "Detector, High resolution, Positron emission tomography (PET), Small animal",
author = "Kisung Lee and Miyaoka, {Robert S.} and Janes, {Marie L.} and Lewellen, {Tom K.}",
year = "2005",
month = "10",
day = "1",
doi = "10.1109/TNS.2005.857627",
language = "English",
volume = "52",
pages = "1428--1433",
journal = "IEEE Transactions on Nuclear Science",
issn = "0018-9499",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
number = "5 I",

}

TY - JOUR

T1 - Detector characteristics of the Micro Crystal Element Scanner (MiCES)

AU - Lee, Kisung

AU - Miyaoka, Robert S.

AU - Janes, Marie L.

AU - Lewellen, Tom K.

PY - 2005/10/1

Y1 - 2005/10/1

N2 - This work reports on the performance characteristics of detector modules for the micro crystal element scanner (MiCES). MiCES is a small animal PET scanner that is designed for imaging mice. It will consist of four rings of 18 detector modules. The crystal array for each detector is a 22 × 22 grid of 0.8 × 0.8 × 10 mm mixed lutetium silicate (MLS) crystals set at 0.9 mm centers and framed for rigidity to compose a 19.94 mm square. A technique to mount multi-element crystal arrays to position sensitive photomultiplier tubes (PSPMTs) is described. The crystal arrays are slightly larger than the dimensions of the anode strips of the PSPMT. Therefore the detector module assembly requires precise alignment of the crystals with the PMT optics for optimum decoding. The crystal arrays are glued to the PSPMT using Histomount (National Diagnostics) glue. It has excellent transmittance at MLS' peak emission wavelength. In addition, tapping down the crystals while the glue was curing was found to significantly improve the amount of light collected. We introduce a semi-automatic boundary segmentation technique to generate look-up-tables used for decoding the crystal of interaction from the collected PSPMT signals. The algorithm is based on valley detection followed by line segment tracking. The software tool provides a more convenient and precise way to create the crystal map for each detector module. Performance characteristics of each of the first 18 detector modules that were assembled and processed with the described techniques are presented. The overall results are an average energy resolution of 23.8% and an average peak-to-valley ratio of 7.6. These averages, however, include modules that had the crystal arrays removed and re-glued multiple times as part of our learning curve of how to optimize the gluing and positioning procedures.

AB - This work reports on the performance characteristics of detector modules for the micro crystal element scanner (MiCES). MiCES is a small animal PET scanner that is designed for imaging mice. It will consist of four rings of 18 detector modules. The crystal array for each detector is a 22 × 22 grid of 0.8 × 0.8 × 10 mm mixed lutetium silicate (MLS) crystals set at 0.9 mm centers and framed for rigidity to compose a 19.94 mm square. A technique to mount multi-element crystal arrays to position sensitive photomultiplier tubes (PSPMTs) is described. The crystal arrays are slightly larger than the dimensions of the anode strips of the PSPMT. Therefore the detector module assembly requires precise alignment of the crystals with the PMT optics for optimum decoding. The crystal arrays are glued to the PSPMT using Histomount (National Diagnostics) glue. It has excellent transmittance at MLS' peak emission wavelength. In addition, tapping down the crystals while the glue was curing was found to significantly improve the amount of light collected. We introduce a semi-automatic boundary segmentation technique to generate look-up-tables used for decoding the crystal of interaction from the collected PSPMT signals. The algorithm is based on valley detection followed by line segment tracking. The software tool provides a more convenient and precise way to create the crystal map for each detector module. Performance characteristics of each of the first 18 detector modules that were assembled and processed with the described techniques are presented. The overall results are an average energy resolution of 23.8% and an average peak-to-valley ratio of 7.6. These averages, however, include modules that had the crystal arrays removed and re-glued multiple times as part of our learning curve of how to optimize the gluing and positioning procedures.

KW - Detector

KW - High resolution

KW - Positron emission tomography (PET)

KW - Small animal

UR - http://www.scopus.com/inward/record.url?scp=29144494837&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=29144494837&partnerID=8YFLogxK

U2 - 10.1109/TNS.2005.857627

DO - 10.1109/TNS.2005.857627

M3 - Article

VL - 52

SP - 1428

EP - 1433

JO - IEEE Transactions on Nuclear Science

JF - IEEE Transactions on Nuclear Science

SN - 0018-9499

IS - 5 I

ER -