Development of a one-stop beam verification system using electronic portal imaging devices for routine quality assurance

Sangwook Lim, Sun Young Ma, Tae Sig Jeung, Byong Yong Yi, Sang Hoon Lee, Suk Lee, Sam Ju Cho, Jinho Choi

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

In this study, a computer-based system for routine quality assurance (QA) of a linear accelerator (linac) was developed by using the dosimetric properties of an amorphous silicon electronic portal imaging device (EPID). An acrylic template phantom was designed such that it could be placed on the EPID and be aligned with the light field of the collimator. After irradiation, portal images obtained from the EPID were transferred in DICOM format to a computer and analyzed using a program we developed. The symmetry, flatness, field size, and congruence of the light and radiation fields of the photon beams from the linac were verified simultaneously. To validate the QA system, the ion chamber and film (X-Omat V2; Kodak, New York, NY) measurements were compared with the EPID measurements obtained in this study. The EPID measurements agreed with the film measurements. Parameters for beams with energies of 6 MV and 15 MV were obtained daily for 1 month using this system. It was found that our QA tool using EPID could substitute for the film test, which is a time-consuming method for routine QA assessment.

Original languageEnglish
Pages (from-to)296-304
Number of pages9
JournalMedical Dosimetry
Volume37
Issue number3
DOIs
Publication statusPublished - 2012 Jan 23

Fingerprint

Equipment and Supplies
Particle Accelerators
Light
Computer Systems
Silicon
Photons
Ions
Radiation

Keywords

  • Daily QA
  • EPID
  • Radiation field

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Development of a one-stop beam verification system using electronic portal imaging devices for routine quality assurance. / Lim, Sangwook; Ma, Sun Young; Jeung, Tae Sig; Yi, Byong Yong; Lee, Sang Hoon; Lee, Suk; Cho, Sam Ju; Choi, Jinho.

In: Medical Dosimetry, Vol. 37, No. 3, 23.01.2012, p. 296-304.

Research output: Contribution to journalArticle

Lim, Sangwook ; Ma, Sun Young ; Jeung, Tae Sig ; Yi, Byong Yong ; Lee, Sang Hoon ; Lee, Suk ; Cho, Sam Ju ; Choi, Jinho. / Development of a one-stop beam verification system using electronic portal imaging devices for routine quality assurance. In: Medical Dosimetry. 2012 ; Vol. 37, No. 3. pp. 296-304.
@article{877364b179c148ccb73b1106d44fab87,
title = "Development of a one-stop beam verification system using electronic portal imaging devices for routine quality assurance",
abstract = "In this study, a computer-based system for routine quality assurance (QA) of a linear accelerator (linac) was developed by using the dosimetric properties of an amorphous silicon electronic portal imaging device (EPID). An acrylic template phantom was designed such that it could be placed on the EPID and be aligned with the light field of the collimator. After irradiation, portal images obtained from the EPID were transferred in DICOM format to a computer and analyzed using a program we developed. The symmetry, flatness, field size, and congruence of the light and radiation fields of the photon beams from the linac were verified simultaneously. To validate the QA system, the ion chamber and film (X-Omat V2; Kodak, New York, NY) measurements were compared with the EPID measurements obtained in this study. The EPID measurements agreed with the film measurements. Parameters for beams with energies of 6 MV and 15 MV were obtained daily for 1 month using this system. It was found that our QA tool using EPID could substitute for the film test, which is a time-consuming method for routine QA assessment.",
keywords = "Daily QA, EPID, Radiation field",
author = "Sangwook Lim and Ma, {Sun Young} and Jeung, {Tae Sig} and Yi, {Byong Yong} and Lee, {Sang Hoon} and Suk Lee and Cho, {Sam Ju} and Jinho Choi",
year = "2012",
month = "1",
day = "23",
doi = "10.1016/j.meddos.2011.11.003",
language = "English",
volume = "37",
pages = "296--304",
journal = "Medical Dosimetry",
issn = "0958-3947",
publisher = "Elsevier Inc.",
number = "3",

}

TY - JOUR

T1 - Development of a one-stop beam verification system using electronic portal imaging devices for routine quality assurance

AU - Lim, Sangwook

AU - Ma, Sun Young

AU - Jeung, Tae Sig

AU - Yi, Byong Yong

AU - Lee, Sang Hoon

AU - Lee, Suk

AU - Cho, Sam Ju

AU - Choi, Jinho

PY - 2012/1/23

Y1 - 2012/1/23

N2 - In this study, a computer-based system for routine quality assurance (QA) of a linear accelerator (linac) was developed by using the dosimetric properties of an amorphous silicon electronic portal imaging device (EPID). An acrylic template phantom was designed such that it could be placed on the EPID and be aligned with the light field of the collimator. After irradiation, portal images obtained from the EPID were transferred in DICOM format to a computer and analyzed using a program we developed. The symmetry, flatness, field size, and congruence of the light and radiation fields of the photon beams from the linac were verified simultaneously. To validate the QA system, the ion chamber and film (X-Omat V2; Kodak, New York, NY) measurements were compared with the EPID measurements obtained in this study. The EPID measurements agreed with the film measurements. Parameters for beams with energies of 6 MV and 15 MV were obtained daily for 1 month using this system. It was found that our QA tool using EPID could substitute for the film test, which is a time-consuming method for routine QA assessment.

AB - In this study, a computer-based system for routine quality assurance (QA) of a linear accelerator (linac) was developed by using the dosimetric properties of an amorphous silicon electronic portal imaging device (EPID). An acrylic template phantom was designed such that it could be placed on the EPID and be aligned with the light field of the collimator. After irradiation, portal images obtained from the EPID were transferred in DICOM format to a computer and analyzed using a program we developed. The symmetry, flatness, field size, and congruence of the light and radiation fields of the photon beams from the linac were verified simultaneously. To validate the QA system, the ion chamber and film (X-Omat V2; Kodak, New York, NY) measurements were compared with the EPID measurements obtained in this study. The EPID measurements agreed with the film measurements. Parameters for beams with energies of 6 MV and 15 MV were obtained daily for 1 month using this system. It was found that our QA tool using EPID could substitute for the film test, which is a time-consuming method for routine QA assessment.

KW - Daily QA

KW - EPID

KW - Radiation field

UR - http://www.scopus.com/inward/record.url?scp=84864779050&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84864779050&partnerID=8YFLogxK

U2 - 10.1016/j.meddos.2011.11.003

DO - 10.1016/j.meddos.2011.11.003

M3 - Article

C2 - 22277157

AN - SCOPUS:84864779050

VL - 37

SP - 296

EP - 304

JO - Medical Dosimetry

JF - Medical Dosimetry

SN - 0958-3947

IS - 3

ER -