Development of a patient dose verification method that uses the transit dose measured with a glass dosimeter

Seonghoon Jeong, Myonggeun Yoon, Dong Wook Kim, Weon Kuu Chung, Mijoo Chung

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

The present study aims to verify patient doses by using a glass dosimeter to measure the radiation penetrating patients during radiation therapy and developing an algorithm based thereon that can simply and accurately calculate the internal doses reserved by the patients. In this study, we developed the algorithm and, as a feasibility study for a clinical test, we verified its accuracy by using a solid water phantom. We developed an algorithm that measured the scatter factor, which represents the correlation between the bottom of the phantoms and the transit-dose measurement unit and calculates the center doses of the phantoms by using the inverse square law and percentage depth dose data. Also, we used a glass dosimeter to measure the transit radiation of 14 patients’ 3D conformal radiation therapy cases and predicted the internal doses of the phantoms by using the results. By comparing these values and those of actual measurements obtained from an ion chamber, we evaluated the accuracy of the developed algorithm. When the developed algorithm was applied for patient dose verification based on the transit doses to 14 patients’ 3D-CRT treatment, as measured using the ion chamber, the average deviation from the measurement values for the phantoms was 0.33% (±0.57%). Based on this result, we confirmed that the algorithm is correct. When the patient doses obtained using a glass dosimeter were verified in the same way, the average deviation was −0.28% (±1.87%). The algorithm for calculating patients’ internal doses based on transit radiation measurements was found to be accurate and is expected to be capable of simply verifying patient doses even when measuring with a glass dosimeter, despite an increased error to a certain degree.

Original languageEnglish
Pages (from-to)948-955
Number of pages8
JournalJournal of the Korean Physical Society
Volume70
Issue number10
DOIs
Publication statusPublished - 2017 May 1

Fingerprint

transit
dosimeters
dosage
glass
ionization chambers
radiation therapy
deviation
radiation measurement
radiation

Keywords

  • Glass dosimeter
  • Inverse square law
  • Patient dose verification
  • Scatter factor
  • Transit dose

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Cite this

Development of a patient dose verification method that uses the transit dose measured with a glass dosimeter. / Jeong, Seonghoon; Yoon, Myonggeun; Kim, Dong Wook; Chung, Weon Kuu; Chung, Mijoo.

In: Journal of the Korean Physical Society, Vol. 70, No. 10, 01.05.2017, p. 948-955.

Research output: Contribution to journalArticle

Jeong, Seonghoon ; Yoon, Myonggeun ; Kim, Dong Wook ; Chung, Weon Kuu ; Chung, Mijoo. / Development of a patient dose verification method that uses the transit dose measured with a glass dosimeter. In: Journal of the Korean Physical Society. 2017 ; Vol. 70, No. 10. pp. 948-955.
@article{7dcb628175cd48338d14e79514f9c669,
title = "Development of a patient dose verification method that uses the transit dose measured with a glass dosimeter",
abstract = "The present study aims to verify patient doses by using a glass dosimeter to measure the radiation penetrating patients during radiation therapy and developing an algorithm based thereon that can simply and accurately calculate the internal doses reserved by the patients. In this study, we developed the algorithm and, as a feasibility study for a clinical test, we verified its accuracy by using a solid water phantom. We developed an algorithm that measured the scatter factor, which represents the correlation between the bottom of the phantoms and the transit-dose measurement unit and calculates the center doses of the phantoms by using the inverse square law and percentage depth dose data. Also, we used a glass dosimeter to measure the transit radiation of 14 patients’ 3D conformal radiation therapy cases and predicted the internal doses of the phantoms by using the results. By comparing these values and those of actual measurements obtained from an ion chamber, we evaluated the accuracy of the developed algorithm. When the developed algorithm was applied for patient dose verification based on the transit doses to 14 patients’ 3D-CRT treatment, as measured using the ion chamber, the average deviation from the measurement values for the phantoms was 0.33{\%} (±0.57{\%}). Based on this result, we confirmed that the algorithm is correct. When the patient doses obtained using a glass dosimeter were verified in the same way, the average deviation was −0.28{\%} (±1.87{\%}). The algorithm for calculating patients’ internal doses based on transit radiation measurements was found to be accurate and is expected to be capable of simply verifying patient doses even when measuring with a glass dosimeter, despite an increased error to a certain degree.",
keywords = "Glass dosimeter, Inverse square law, Patient dose verification, Scatter factor, Transit dose",
author = "Seonghoon Jeong and Myonggeun Yoon and Kim, {Dong Wook} and Chung, {Weon Kuu} and Mijoo Chung",
year = "2017",
month = "5",
day = "1",
doi = "10.3938/jkps.70.948",
language = "English",
volume = "70",
pages = "948--955",
journal = "Journal of the Korean Physical Society",
issn = "0374-4884",
publisher = "Korean Physical Society",
number = "10",

}

TY - JOUR

T1 - Development of a patient dose verification method that uses the transit dose measured with a glass dosimeter

AU - Jeong, Seonghoon

AU - Yoon, Myonggeun

AU - Kim, Dong Wook

AU - Chung, Weon Kuu

AU - Chung, Mijoo

PY - 2017/5/1

Y1 - 2017/5/1

N2 - The present study aims to verify patient doses by using a glass dosimeter to measure the radiation penetrating patients during radiation therapy and developing an algorithm based thereon that can simply and accurately calculate the internal doses reserved by the patients. In this study, we developed the algorithm and, as a feasibility study for a clinical test, we verified its accuracy by using a solid water phantom. We developed an algorithm that measured the scatter factor, which represents the correlation between the bottom of the phantoms and the transit-dose measurement unit and calculates the center doses of the phantoms by using the inverse square law and percentage depth dose data. Also, we used a glass dosimeter to measure the transit radiation of 14 patients’ 3D conformal radiation therapy cases and predicted the internal doses of the phantoms by using the results. By comparing these values and those of actual measurements obtained from an ion chamber, we evaluated the accuracy of the developed algorithm. When the developed algorithm was applied for patient dose verification based on the transit doses to 14 patients’ 3D-CRT treatment, as measured using the ion chamber, the average deviation from the measurement values for the phantoms was 0.33% (±0.57%). Based on this result, we confirmed that the algorithm is correct. When the patient doses obtained using a glass dosimeter were verified in the same way, the average deviation was −0.28% (±1.87%). The algorithm for calculating patients’ internal doses based on transit radiation measurements was found to be accurate and is expected to be capable of simply verifying patient doses even when measuring with a glass dosimeter, despite an increased error to a certain degree.

AB - The present study aims to verify patient doses by using a glass dosimeter to measure the radiation penetrating patients during radiation therapy and developing an algorithm based thereon that can simply and accurately calculate the internal doses reserved by the patients. In this study, we developed the algorithm and, as a feasibility study for a clinical test, we verified its accuracy by using a solid water phantom. We developed an algorithm that measured the scatter factor, which represents the correlation between the bottom of the phantoms and the transit-dose measurement unit and calculates the center doses of the phantoms by using the inverse square law and percentage depth dose data. Also, we used a glass dosimeter to measure the transit radiation of 14 patients’ 3D conformal radiation therapy cases and predicted the internal doses of the phantoms by using the results. By comparing these values and those of actual measurements obtained from an ion chamber, we evaluated the accuracy of the developed algorithm. When the developed algorithm was applied for patient dose verification based on the transit doses to 14 patients’ 3D-CRT treatment, as measured using the ion chamber, the average deviation from the measurement values for the phantoms was 0.33% (±0.57%). Based on this result, we confirmed that the algorithm is correct. When the patient doses obtained using a glass dosimeter were verified in the same way, the average deviation was −0.28% (±1.87%). The algorithm for calculating patients’ internal doses based on transit radiation measurements was found to be accurate and is expected to be capable of simply verifying patient doses even when measuring with a glass dosimeter, despite an increased error to a certain degree.

KW - Glass dosimeter

KW - Inverse square law

KW - Patient dose verification

KW - Scatter factor

KW - Transit dose

UR - http://www.scopus.com/inward/record.url?scp=85019999006&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85019999006&partnerID=8YFLogxK

U2 - 10.3938/jkps.70.948

DO - 10.3938/jkps.70.948

M3 - Article

AN - SCOPUS:85019999006

VL - 70

SP - 948

EP - 955

JO - Journal of the Korean Physical Society

JF - Journal of the Korean Physical Society

SN - 0374-4884

IS - 10

ER -