Development of a program for toric intraocular lens calculation considering posterior corneal astigmatism, incision-induced posterior corneal astigmatism, and effective lens position

Youngsub Eom, Dongok Ryu, Dae Wook Kim, Seul Ki Yang, Jong Suk Song, Sug Whan Kim, Hyo Myung Kim

    Research output: Contribution to journalArticlepeer-review

    9 Citations (Scopus)

    Abstract

    Background: To evaluate the toric intraocular lens (IOL) calculation considering posterior corneal astigmatism, incision-induced posterior corneal astigmatism, and effective lens position (ELP). Methods: Two thousand samples of corneal parameters with keratometric astigmatism ≥ 1.0 D were obtained using bootstrap methods. The probability distributions for incision-induced keratometric and posterior corneal astigmatisms, as well as ELP were estimated from the literature review. The predicted residual astigmatism error using method D with an IOL add power calculator (IAPC) was compared with those derived using methods A, B, and C through Monte-Carlo simulation. Method A considered the keratometric astigmatism and incision-induced keratometric astigmatism, method B considered posterior corneal astigmatism in addition to the A method, method C considered incision-induced posterior corneal astigmatism in addition to the B method, and method D considered ELP in addition to the C method. To verify the IAPC used in this study, the predicted toric IOL cylinder power and its axis using the IAPC were compared with ray-tracing simulation results. Results: The median magnitude of the predicted residual astigmatism error using method D (0.25 diopters [D]) was smaller than that derived using methods A (0.42 D), B (0.38 D), and C (0.28 D) respectively. Linear regression analysis indicated that the predicted toric IOL cylinder power and its axis had excellent goodness-of-fit between the IAPC and ray-tracing simulation. Conclusions: The IAPC is a simple but accurate method for predicting the toric IOL cylinder power and its axis considering posterior corneal astigmatism, incision-induced posterior corneal astigmatism, and ELP.

    Original languageEnglish
    Pages (from-to)1977-1986
    Number of pages10
    JournalGraefe's Archive for Clinical and Experimental Ophthalmology
    Volume254
    Issue number10
    DOIs
    Publication statusPublished - 2016 Oct 1

    Keywords

    • Effective lens position
    • Posterior corneal astigmatism, Incision-induced astigmatism
    • Toric intraocular lens

    ASJC Scopus subject areas

    • Ophthalmology
    • Sensory Systems
    • Cellular and Molecular Neuroscience