Effect of radiation scattering on dose uniformity in open and closed cell culture vessels

Myonggeun Yoon, Sung Yong Park, Jungwook Shin, Jungwon Kwak, Junghun Park, Dongho Shin, Soah Park, Se Byeong Lee, Doo Hyun Lee, Kyung Hwan Shin, Hong Ryull Pyo, Joo Young Kim, Kwan Ho Cho

Research output: Contribution to journalArticle

Abstract

Purpose: Dose uniformity in cell culture vessels such as Petri dishes and anoxic irradiation chambers is very important in radiobiological work as dose uniformity affects cell survival probabilities. In this study, we investigated X-ray dose inhomogeneity, caused by scattering, in typical culture vessels. Materials and methods: Three different cubic cell culture vessels, with side lengths of 10 cm, 15 cm and 20 cm, were designed and irradiated by X-rays of 6 MV and 15 MV at a source-to-surface distance (SSD) of 100 cm using a Varian 2100CD linear accelerator. Results: The relative X-ray dose distribution in a cell culture vessel depended strongly on whether the vessel had a lid. The percentage of the cell culture surface with the dose differing by more than 10% from the mean value of the dose was 43.4% in lidless vessels and 9.7% in lidded vessels. Conclusions: In radiobiological work, X-ray dose inhomogeneity within a cell culture vessel is not negligible and the placement of cells in the vessel should be carefully considered.

Original languageEnglish
Pages (from-to)561-566
Number of pages6
JournalInternational Journal of Radiation Biology
Volume83
Issue number8
DOIs
Publication statusPublished - 2007 Jul 12
Externally publishedYes

Fingerprint

Radiation Scattering
Cell culture
Dosimetry
vessels
cell culture
Cell Culture Techniques
Scattering
Radiation
dosage
X-Rays
radiation
X rays
scattering
X-radiation
Particle Accelerators
Linear accelerators
inhomogeneity
x rays
lids
Cell Survival

Keywords

  • Cell culture vessel
  • Homogeneity
  • Oxygen enhancement ratio
  • Radiobiology
  • Scattering

ASJC Scopus subject areas

  • Agricultural and Biological Sciences (miscellaneous)
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology
  • Nuclear Energy and Engineering
  • Radiation

Cite this

Effect of radiation scattering on dose uniformity in open and closed cell culture vessels. / Yoon, Myonggeun; Park, Sung Yong; Shin, Jungwook; Kwak, Jungwon; Park, Junghun; Shin, Dongho; Park, Soah; Lee, Se Byeong; Lee, Doo Hyun; Shin, Kyung Hwan; Pyo, Hong Ryull; Kim, Joo Young; Cho, Kwan Ho.

In: International Journal of Radiation Biology, Vol. 83, No. 8, 12.07.2007, p. 561-566.

Research output: Contribution to journalArticle

Yoon, M, Park, SY, Shin, J, Kwak, J, Park, J, Shin, D, Park, S, Lee, SB, Lee, DH, Shin, KH, Pyo, HR, Kim, JY & Cho, KH 2007, 'Effect of radiation scattering on dose uniformity in open and closed cell culture vessels', International Journal of Radiation Biology, vol. 83, no. 8, pp. 561-566. https://doi.org/10.1080/09553000701383178
Yoon, Myonggeun ; Park, Sung Yong ; Shin, Jungwook ; Kwak, Jungwon ; Park, Junghun ; Shin, Dongho ; Park, Soah ; Lee, Se Byeong ; Lee, Doo Hyun ; Shin, Kyung Hwan ; Pyo, Hong Ryull ; Kim, Joo Young ; Cho, Kwan Ho. / Effect of radiation scattering on dose uniformity in open and closed cell culture vessels. In: International Journal of Radiation Biology. 2007 ; Vol. 83, No. 8. pp. 561-566.
@article{b7623b34c9b646f5b8736fbe989a20c3,
title = "Effect of radiation scattering on dose uniformity in open and closed cell culture vessels",
abstract = "Purpose: Dose uniformity in cell culture vessels such as Petri dishes and anoxic irradiation chambers is very important in radiobiological work as dose uniformity affects cell survival probabilities. In this study, we investigated X-ray dose inhomogeneity, caused by scattering, in typical culture vessels. Materials and methods: Three different cubic cell culture vessels, with side lengths of 10 cm, 15 cm and 20 cm, were designed and irradiated by X-rays of 6 MV and 15 MV at a source-to-surface distance (SSD) of 100 cm using a Varian 2100CD linear accelerator. Results: The relative X-ray dose distribution in a cell culture vessel depended strongly on whether the vessel had a lid. The percentage of the cell culture surface with the dose differing by more than 10{\%} from the mean value of the dose was 43.4{\%} in lidless vessels and 9.7{\%} in lidded vessels. Conclusions: In radiobiological work, X-ray dose inhomogeneity within a cell culture vessel is not negligible and the placement of cells in the vessel should be carefully considered.",
keywords = "Cell culture vessel, Homogeneity, Oxygen enhancement ratio, Radiobiology, Scattering",
author = "Myonggeun Yoon and Park, {Sung Yong} and Jungwook Shin and Jungwon Kwak and Junghun Park and Dongho Shin and Soah Park and Lee, {Se Byeong} and Lee, {Doo Hyun} and Shin, {Kyung Hwan} and Pyo, {Hong Ryull} and Kim, {Joo Young} and Cho, {Kwan Ho}",
year = "2007",
month = "7",
day = "12",
doi = "10.1080/09553000701383178",
language = "English",
volume = "83",
pages = "561--566",
journal = "International Journal of Radiation Biology",
issn = "0955-3002",
publisher = "Informa Healthcare",
number = "8",

}

TY - JOUR

T1 - Effect of radiation scattering on dose uniformity in open and closed cell culture vessels

AU - Yoon, Myonggeun

AU - Park, Sung Yong

AU - Shin, Jungwook

AU - Kwak, Jungwon

AU - Park, Junghun

AU - Shin, Dongho

AU - Park, Soah

AU - Lee, Se Byeong

AU - Lee, Doo Hyun

AU - Shin, Kyung Hwan

AU - Pyo, Hong Ryull

AU - Kim, Joo Young

AU - Cho, Kwan Ho

PY - 2007/7/12

Y1 - 2007/7/12

N2 - Purpose: Dose uniformity in cell culture vessels such as Petri dishes and anoxic irradiation chambers is very important in radiobiological work as dose uniformity affects cell survival probabilities. In this study, we investigated X-ray dose inhomogeneity, caused by scattering, in typical culture vessels. Materials and methods: Three different cubic cell culture vessels, with side lengths of 10 cm, 15 cm and 20 cm, were designed and irradiated by X-rays of 6 MV and 15 MV at a source-to-surface distance (SSD) of 100 cm using a Varian 2100CD linear accelerator. Results: The relative X-ray dose distribution in a cell culture vessel depended strongly on whether the vessel had a lid. The percentage of the cell culture surface with the dose differing by more than 10% from the mean value of the dose was 43.4% in lidless vessels and 9.7% in lidded vessels. Conclusions: In radiobiological work, X-ray dose inhomogeneity within a cell culture vessel is not negligible and the placement of cells in the vessel should be carefully considered.

AB - Purpose: Dose uniformity in cell culture vessels such as Petri dishes and anoxic irradiation chambers is very important in radiobiological work as dose uniformity affects cell survival probabilities. In this study, we investigated X-ray dose inhomogeneity, caused by scattering, in typical culture vessels. Materials and methods: Three different cubic cell culture vessels, with side lengths of 10 cm, 15 cm and 20 cm, were designed and irradiated by X-rays of 6 MV and 15 MV at a source-to-surface distance (SSD) of 100 cm using a Varian 2100CD linear accelerator. Results: The relative X-ray dose distribution in a cell culture vessel depended strongly on whether the vessel had a lid. The percentage of the cell culture surface with the dose differing by more than 10% from the mean value of the dose was 43.4% in lidless vessels and 9.7% in lidded vessels. Conclusions: In radiobiological work, X-ray dose inhomogeneity within a cell culture vessel is not negligible and the placement of cells in the vessel should be carefully considered.

KW - Cell culture vessel

KW - Homogeneity

KW - Oxygen enhancement ratio

KW - Radiobiology

KW - Scattering

UR - http://www.scopus.com/inward/record.url?scp=34447116365&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34447116365&partnerID=8YFLogxK

U2 - 10.1080/09553000701383178

DO - 10.1080/09553000701383178

M3 - Article

C2 - 17613129

AN - SCOPUS:34447116365

VL - 83

SP - 561

EP - 566

JO - International Journal of Radiation Biology

JF - International Journal of Radiation Biology

SN - 0955-3002

IS - 8