Abstract
The radiation exposure dose must be optimised because the hazard resulting from an interventional radiology procedure is long term depending on the patient. The aim of this study was to measure the radiation doses received by the patients and medical staff during endoscopic retrograde cholangiopancreatography (ERCP) procedures. Data were collected during 126 ERCP procedures, including the dose-area product (DAP), entrance dose (ED), effective dose (E), fluoroscopy time (T) and number of digital radiographs (F). The medical staff members each wore a personal thermoluminescence dosemeter to monitor exposure during ERCP procedures. The mean DAP, ED, E and T were 47.06 Gy cm2, 196.06 mGy, 8.93 mSv, 7.65 min and 9.21 images, respectively. The mean dose to the staff was 0.175 mSv and that to the assistant was 0.069 mSv. The dose to the medical staff was minimal when appropriate protective measures were used. The large variation in the patient doses must be further investigated.