Evaluation of the dosimetric accuracy for a couch-based tracking system (CBTS)

Kyung Hwan Chang, Suk Lee, Kwang Hyeon Kim, Jang Bo Shim, Dae-Sik Yang, Young Je Park, Won Sup Yoon, Chul Yong Kim, Yuanjie Cao

Research output: Contribution to journalArticle

Abstract

In this study, the geometric and dosimetric accuracy of an in-house-developed couch-based tracking system (CBTS) was investigated using both film and in-house-developed polymer gel dosimeters. We evaluated the 1D and the 2D motion accuracies of our couch system by using Gafchromic EBT film. For the 1D test, the couch system was moved 5, 10, and 20 mm in the X, Y, and Z directions, respectively. Meanwhile, for the 2D test, it was moved along the XY, YZ, and ZX directions. We compared the profiles, full widths at half maximum (FWHMs), and penumbras between the static and the tracking fields. For the 3D test, we quantitatively compared the dose distribution between the static and the tracking fields by using the polymer gel dosimeter when it was simultaneously moved in the XYZ directions. We confirmed that the film was moved according to motion amplitudes of 5, 10, and 20 mm in the X, Y, and Z directions, respectively, in the 1D and 2D motion tests. The value of the FWHM of the static field and the three tracking fields were 51.88, 53.28, 57.67, and 64.43 mm, respectively. Two types of penumbras became wider with increasing amplitudes compared to the static field. For the 3D test, the dose distribution of the XYZ tracking field was qualitatively larger than that of the static field. We conclude that this CBTS has the potential for pre-clinical applications in adaptive radiation therapy.

Original languageEnglish
Pages (from-to)241-247
Number of pages7
JournalJournal of the Korean Physical Society
Volume69
Issue number2
DOIs
Publication statusPublished - 2016 Jul 1

Fingerprint

couches
evaluation
penumbras
dosimeters
geometric accuracy
gels
dosage
polymers
radiation therapy
profiles

Keywords

  • Dosimetric accuracy
  • Gel dosimeter
  • Patient motion
  • Robotic couch
  • Tumor tracking

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Cite this

Evaluation of the dosimetric accuracy for a couch-based tracking system (CBTS). / Chang, Kyung Hwan; Lee, Suk; Kim, Kwang Hyeon; Shim, Jang Bo; Yang, Dae-Sik; Park, Young Je; Yoon, Won Sup; Kim, Chul Yong; Cao, Yuanjie.

In: Journal of the Korean Physical Society, Vol. 69, No. 2, 01.07.2016, p. 241-247.

Research output: Contribution to journalArticle

Chang, Kyung Hwan ; Lee, Suk ; Kim, Kwang Hyeon ; Shim, Jang Bo ; Yang, Dae-Sik ; Park, Young Je ; Yoon, Won Sup ; Kim, Chul Yong ; Cao, Yuanjie. / Evaluation of the dosimetric accuracy for a couch-based tracking system (CBTS). In: Journal of the Korean Physical Society. 2016 ; Vol. 69, No. 2. pp. 241-247.
@article{a31609ac82404423a97c9ffdbf8f1d61,
title = "Evaluation of the dosimetric accuracy for a couch-based tracking system (CBTS)",
abstract = "In this study, the geometric and dosimetric accuracy of an in-house-developed couch-based tracking system (CBTS) was investigated using both film and in-house-developed polymer gel dosimeters. We evaluated the 1D and the 2D motion accuracies of our couch system by using Gafchromic EBT film. For the 1D test, the couch system was moved 5, 10, and 20 mm in the X, Y, and Z directions, respectively. Meanwhile, for the 2D test, it was moved along the XY, YZ, and ZX directions. We compared the profiles, full widths at half maximum (FWHMs), and penumbras between the static and the tracking fields. For the 3D test, we quantitatively compared the dose distribution between the static and the tracking fields by using the polymer gel dosimeter when it was simultaneously moved in the XYZ directions. We confirmed that the film was moved according to motion amplitudes of 5, 10, and 20 mm in the X, Y, and Z directions, respectively, in the 1D and 2D motion tests. The value of the FWHM of the static field and the three tracking fields were 51.88, 53.28, 57.67, and 64.43 mm, respectively. Two types of penumbras became wider with increasing amplitudes compared to the static field. For the 3D test, the dose distribution of the XYZ tracking field was qualitatively larger than that of the static field. We conclude that this CBTS has the potential for pre-clinical applications in adaptive radiation therapy.",
keywords = "Dosimetric accuracy, Gel dosimeter, Patient motion, Robotic couch, Tumor tracking",
author = "Chang, {Kyung Hwan} and Suk Lee and Kim, {Kwang Hyeon} and Shim, {Jang Bo} and Dae-Sik Yang and Park, {Young Je} and Yoon, {Won Sup} and Kim, {Chul Yong} and Yuanjie Cao",
year = "2016",
month = "7",
day = "1",
doi = "10.3938/jkps.69.241",
language = "English",
volume = "69",
pages = "241--247",
journal = "Journal of the Korean Physical Society",
issn = "0374-4884",
publisher = "Korean Physical Society",
number = "2",

}

TY - JOUR

T1 - Evaluation of the dosimetric accuracy for a couch-based tracking system (CBTS)

AU - Chang, Kyung Hwan

AU - Lee, Suk

AU - Kim, Kwang Hyeon

AU - Shim, Jang Bo

AU - Yang, Dae-Sik

AU - Park, Young Je

AU - Yoon, Won Sup

AU - Kim, Chul Yong

AU - Cao, Yuanjie

PY - 2016/7/1

Y1 - 2016/7/1

N2 - In this study, the geometric and dosimetric accuracy of an in-house-developed couch-based tracking system (CBTS) was investigated using both film and in-house-developed polymer gel dosimeters. We evaluated the 1D and the 2D motion accuracies of our couch system by using Gafchromic EBT film. For the 1D test, the couch system was moved 5, 10, and 20 mm in the X, Y, and Z directions, respectively. Meanwhile, for the 2D test, it was moved along the XY, YZ, and ZX directions. We compared the profiles, full widths at half maximum (FWHMs), and penumbras between the static and the tracking fields. For the 3D test, we quantitatively compared the dose distribution between the static and the tracking fields by using the polymer gel dosimeter when it was simultaneously moved in the XYZ directions. We confirmed that the film was moved according to motion amplitudes of 5, 10, and 20 mm in the X, Y, and Z directions, respectively, in the 1D and 2D motion tests. The value of the FWHM of the static field and the three tracking fields were 51.88, 53.28, 57.67, and 64.43 mm, respectively. Two types of penumbras became wider with increasing amplitudes compared to the static field. For the 3D test, the dose distribution of the XYZ tracking field was qualitatively larger than that of the static field. We conclude that this CBTS has the potential for pre-clinical applications in adaptive radiation therapy.

AB - In this study, the geometric and dosimetric accuracy of an in-house-developed couch-based tracking system (CBTS) was investigated using both film and in-house-developed polymer gel dosimeters. We evaluated the 1D and the 2D motion accuracies of our couch system by using Gafchromic EBT film. For the 1D test, the couch system was moved 5, 10, and 20 mm in the X, Y, and Z directions, respectively. Meanwhile, for the 2D test, it was moved along the XY, YZ, and ZX directions. We compared the profiles, full widths at half maximum (FWHMs), and penumbras between the static and the tracking fields. For the 3D test, we quantitatively compared the dose distribution between the static and the tracking fields by using the polymer gel dosimeter when it was simultaneously moved in the XYZ directions. We confirmed that the film was moved according to motion amplitudes of 5, 10, and 20 mm in the X, Y, and Z directions, respectively, in the 1D and 2D motion tests. The value of the FWHM of the static field and the three tracking fields were 51.88, 53.28, 57.67, and 64.43 mm, respectively. Two types of penumbras became wider with increasing amplitudes compared to the static field. For the 3D test, the dose distribution of the XYZ tracking field was qualitatively larger than that of the static field. We conclude that this CBTS has the potential for pre-clinical applications in adaptive radiation therapy.

KW - Dosimetric accuracy

KW - Gel dosimeter

KW - Patient motion

KW - Robotic couch

KW - Tumor tracking

UR - http://www.scopus.com/inward/record.url?scp=84979578007&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84979578007&partnerID=8YFLogxK

U2 - 10.3938/jkps.69.241

DO - 10.3938/jkps.69.241

M3 - Article

VL - 69

SP - 241

EP - 247

JO - Journal of the Korean Physical Society

JF - Journal of the Korean Physical Society

SN - 0374-4884

IS - 2

ER -