Experimental evaluation of a multi-pinhole collimator for a small organ by using a small-field-of-view gamma camera

Jaekeon Bae, Seungbin Bae, Young Jun Jung, Kisung Lee, Yongkwon Kim, Jinhun Joung, Kyeong Min Kim, Hee Joung Kim

Research output: Contribution to journalArticle

Abstract

The aim of this study is to design and evaluate a multi-pinhole (MP) collimator for a gamma imaging system that requires a high sensitivity, organ-specific, and small footprint. To ensure these requirements, we designed an eight-hole collimator that can be integrated into a small field-of-view gamma camera for imaging the thyroid or relatively sized organs. Each pinhole was designed to have a cylindrical shape with a 2-mm diameter. Experiments were performed with both a two-sphere phantom and a four-rod phantom. An image reconstruction based on the maximum likelihood expectation maximization with the distance-driven method was used for obtaining a 3-dimensional image. For improving the uniformity of the reconstruction image, we modeled the sensitivity of the cylindrical pinhole by calculating the area of the overlapped circle. The results show that the full width at half maximum values of the two-sphere phantom and the four-rod phantom were 7.56 mm (5-mm-diameter source) and 6.84 mm (5-mm-diameter rod), respectively. The scanning time can be reduced by up to 20 minutes in small-organ applications by using developed MP collimator. Thus, the results indicate that the proposed MP collimator is suitable for a fast scan time, as well as for organ-specific and small-footprint applications.

Original languageEnglish
Pages (from-to)416-423
Number of pages8
JournalJournal of the Korean Physical Society
Volume70
Issue number4
DOIs
Publication statusPublished - 2017 Feb 1

Fingerprint

pinholes
collimators
organs
field of view
cameras
evaluation
rods
footprints
image reconstruction
sensitivity
requirements
scanning

Keywords

  • Collimator modeling
  • Experimental data
  • Image reconstruction
  • Multi-pinhole collimator
  • Single-photon emission computed tomography

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Cite this

Experimental evaluation of a multi-pinhole collimator for a small organ by using a small-field-of-view gamma camera. / Bae, Jaekeon; Bae, Seungbin; Jung, Young Jun; Lee, Kisung; Kim, Yongkwon; Joung, Jinhun; Kim, Kyeong Min; Kim, Hee Joung.

In: Journal of the Korean Physical Society, Vol. 70, No. 4, 01.02.2017, p. 416-423.

Research output: Contribution to journalArticle

Bae, Jaekeon ; Bae, Seungbin ; Jung, Young Jun ; Lee, Kisung ; Kim, Yongkwon ; Joung, Jinhun ; Kim, Kyeong Min ; Kim, Hee Joung. / Experimental evaluation of a multi-pinhole collimator for a small organ by using a small-field-of-view gamma camera. In: Journal of the Korean Physical Society. 2017 ; Vol. 70, No. 4. pp. 416-423.
@article{0a9650adaafb47c1826cf5926d363828,
title = "Experimental evaluation of a multi-pinhole collimator for a small organ by using a small-field-of-view gamma camera",
abstract = "The aim of this study is to design and evaluate a multi-pinhole (MP) collimator for a gamma imaging system that requires a high sensitivity, organ-specific, and small footprint. To ensure these requirements, we designed an eight-hole collimator that can be integrated into a small field-of-view gamma camera for imaging the thyroid or relatively sized organs. Each pinhole was designed to have a cylindrical shape with a 2-mm diameter. Experiments were performed with both a two-sphere phantom and a four-rod phantom. An image reconstruction based on the maximum likelihood expectation maximization with the distance-driven method was used for obtaining a 3-dimensional image. For improving the uniformity of the reconstruction image, we modeled the sensitivity of the cylindrical pinhole by calculating the area of the overlapped circle. The results show that the full width at half maximum values of the two-sphere phantom and the four-rod phantom were 7.56 mm (5-mm-diameter source) and 6.84 mm (5-mm-diameter rod), respectively. The scanning time can be reduced by up to 20 minutes in small-organ applications by using developed MP collimator. Thus, the results indicate that the proposed MP collimator is suitable for a fast scan time, as well as for organ-specific and small-footprint applications.",
keywords = "Collimator modeling, Experimental data, Image reconstruction, Multi-pinhole collimator, Single-photon emission computed tomography",
author = "Jaekeon Bae and Seungbin Bae and Jung, {Young Jun} and Kisung Lee and Yongkwon Kim and Jinhun Joung and Kim, {Kyeong Min} and Kim, {Hee Joung}",
year = "2017",
month = "2",
day = "1",
doi = "10.3938/jkps.70.416",
language = "English",
volume = "70",
pages = "416--423",
journal = "Journal of the Korean Physical Society",
issn = "0374-4884",
publisher = "Korean Physical Society",
number = "4",

}

TY - JOUR

T1 - Experimental evaluation of a multi-pinhole collimator for a small organ by using a small-field-of-view gamma camera

AU - Bae, Jaekeon

AU - Bae, Seungbin

AU - Jung, Young Jun

AU - Lee, Kisung

AU - Kim, Yongkwon

AU - Joung, Jinhun

AU - Kim, Kyeong Min

AU - Kim, Hee Joung

PY - 2017/2/1

Y1 - 2017/2/1

N2 - The aim of this study is to design and evaluate a multi-pinhole (MP) collimator for a gamma imaging system that requires a high sensitivity, organ-specific, and small footprint. To ensure these requirements, we designed an eight-hole collimator that can be integrated into a small field-of-view gamma camera for imaging the thyroid or relatively sized organs. Each pinhole was designed to have a cylindrical shape with a 2-mm diameter. Experiments were performed with both a two-sphere phantom and a four-rod phantom. An image reconstruction based on the maximum likelihood expectation maximization with the distance-driven method was used for obtaining a 3-dimensional image. For improving the uniformity of the reconstruction image, we modeled the sensitivity of the cylindrical pinhole by calculating the area of the overlapped circle. The results show that the full width at half maximum values of the two-sphere phantom and the four-rod phantom were 7.56 mm (5-mm-diameter source) and 6.84 mm (5-mm-diameter rod), respectively. The scanning time can be reduced by up to 20 minutes in small-organ applications by using developed MP collimator. Thus, the results indicate that the proposed MP collimator is suitable for a fast scan time, as well as for organ-specific and small-footprint applications.

AB - The aim of this study is to design and evaluate a multi-pinhole (MP) collimator for a gamma imaging system that requires a high sensitivity, organ-specific, and small footprint. To ensure these requirements, we designed an eight-hole collimator that can be integrated into a small field-of-view gamma camera for imaging the thyroid or relatively sized organs. Each pinhole was designed to have a cylindrical shape with a 2-mm diameter. Experiments were performed with both a two-sphere phantom and a four-rod phantom. An image reconstruction based on the maximum likelihood expectation maximization with the distance-driven method was used for obtaining a 3-dimensional image. For improving the uniformity of the reconstruction image, we modeled the sensitivity of the cylindrical pinhole by calculating the area of the overlapped circle. The results show that the full width at half maximum values of the two-sphere phantom and the four-rod phantom were 7.56 mm (5-mm-diameter source) and 6.84 mm (5-mm-diameter rod), respectively. The scanning time can be reduced by up to 20 minutes in small-organ applications by using developed MP collimator. Thus, the results indicate that the proposed MP collimator is suitable for a fast scan time, as well as for organ-specific and small-footprint applications.

KW - Collimator modeling

KW - Experimental data

KW - Image reconstruction

KW - Multi-pinhole collimator

KW - Single-photon emission computed tomography

UR - http://www.scopus.com/inward/record.url?scp=85014506954&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85014506954&partnerID=8YFLogxK

U2 - 10.3938/jkps.70.416

DO - 10.3938/jkps.70.416

M3 - Article

AN - SCOPUS:85014506954

VL - 70

SP - 416

EP - 423

JO - Journal of the Korean Physical Society

JF - Journal of the Korean Physical Society

SN - 0374-4884

IS - 4

ER -