Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

Seong S. Baek, Ji J. Chung, Jaeman Son, Myonggeun Yoon

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

Purpose: The aim of this study is to evaluate the ability of transit dosimetry using commercial treatment planning system (TPS) and an electronic portal imaging device (EPID) with simple calibration method to verify the beam delivery based on detection of large errors in treatment room. Methods and materials: Twenty four fields of intensity modulated radiotherapy (IMRT) plans were selected from four lung cancer patients and used in the irradiation of an anthropomorphic phantom. The proposed method was evaluated by comparing the calculated dose map from TPS and EPID measurement on the same plane using a gamma index method with a 3% dose and 3 mm distance-to-dose agreement tolerance limit. Results: In a simulation using a homogeneous plastic water phantom, performed to verify the effectiveness of the proposed method, the average passing rate of the transit dose based on gamma index was high enough, averaging 94.2% when there was no error during beam delivery. The passing rate of the transit dose for 24 IMRT fields was lower with the anthropomorphic phantom, averaging 86.8% ± 3.8%, a reduction partially due to the inaccuracy of TPS calculations for inhomogeneity. Compared with the TPS, the absolute value of the transit dose at the beam center differed by -0.38% ± 2.1%. The simulation study indicated that the passing rate of the gamma index was significantly reduced, to less than 40%, when a wrong field was erroneously irradiated to patient in the treatment room. Conclusions: This feasibility study suggested that transit dosimetry based on the calculation with commercial TPS and EPID measurement with simple calibration can provide information about large errors for treatment beam delivery.

Original languageEnglish
Article number273
JournalRadiation Oncology
Volume9
Issue number1
DOIs
Publication statusPublished - 2014 Dec 4

Fingerprint

Feasibility Studies
Equipment and Supplies
Portal System
Intensity-Modulated Radiotherapy
Therapeutics
Calibration
Plastics
Lung Neoplasms
Water

Keywords

  • EPID
  • Gamma index
  • Intensity modulated radiation therapy
  • Transit dose

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

Cite this

Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry. / Baek, Seong S.; Chung, Ji J.; Son, Jaeman; Yoon, Myonggeun.

In: Radiation Oncology, Vol. 9, No. 1, 273, 04.12.2014.

Research output: Contribution to journalArticle

@article{cfe0c7f0f2b14aae828f69573578e40d,
title = "Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry",
abstract = "Purpose: The aim of this study is to evaluate the ability of transit dosimetry using commercial treatment planning system (TPS) and an electronic portal imaging device (EPID) with simple calibration method to verify the beam delivery based on detection of large errors in treatment room. Methods and materials: Twenty four fields of intensity modulated radiotherapy (IMRT) plans were selected from four lung cancer patients and used in the irradiation of an anthropomorphic phantom. The proposed method was evaluated by comparing the calculated dose map from TPS and EPID measurement on the same plane using a gamma index method with a 3{\%} dose and 3 mm distance-to-dose agreement tolerance limit. Results: In a simulation using a homogeneous plastic water phantom, performed to verify the effectiveness of the proposed method, the average passing rate of the transit dose based on gamma index was high enough, averaging 94.2{\%} when there was no error during beam delivery. The passing rate of the transit dose for 24 IMRT fields was lower with the anthropomorphic phantom, averaging 86.8{\%} ± 3.8{\%}, a reduction partially due to the inaccuracy of TPS calculations for inhomogeneity. Compared with the TPS, the absolute value of the transit dose at the beam center differed by -0.38{\%} ± 2.1{\%}. The simulation study indicated that the passing rate of the gamma index was significantly reduced, to less than 40{\%}, when a wrong field was erroneously irradiated to patient in the treatment room. Conclusions: This feasibility study suggested that transit dosimetry based on the calculation with commercial TPS and EPID measurement with simple calibration can provide information about large errors for treatment beam delivery.",
keywords = "EPID, Gamma index, Intensity modulated radiation therapy, Transit dose",
author = "Baek, {Seong S.} and Chung, {Ji J.} and Jaeman Son and Myonggeun Yoon",
year = "2014",
month = "12",
day = "4",
doi = "10.1186/s13014-014-0273-8",
language = "English",
volume = "9",
journal = "Radiation Oncology",
issn = "1748-717X",
publisher = "BioMed Central",
number = "1",

}

TY - JOUR

T1 - Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

AU - Baek, Seong S.

AU - Chung, Ji J.

AU - Son, Jaeman

AU - Yoon, Myonggeun

PY - 2014/12/4

Y1 - 2014/12/4

N2 - Purpose: The aim of this study is to evaluate the ability of transit dosimetry using commercial treatment planning system (TPS) and an electronic portal imaging device (EPID) with simple calibration method to verify the beam delivery based on detection of large errors in treatment room. Methods and materials: Twenty four fields of intensity modulated radiotherapy (IMRT) plans were selected from four lung cancer patients and used in the irradiation of an anthropomorphic phantom. The proposed method was evaluated by comparing the calculated dose map from TPS and EPID measurement on the same plane using a gamma index method with a 3% dose and 3 mm distance-to-dose agreement tolerance limit. Results: In a simulation using a homogeneous plastic water phantom, performed to verify the effectiveness of the proposed method, the average passing rate of the transit dose based on gamma index was high enough, averaging 94.2% when there was no error during beam delivery. The passing rate of the transit dose for 24 IMRT fields was lower with the anthropomorphic phantom, averaging 86.8% ± 3.8%, a reduction partially due to the inaccuracy of TPS calculations for inhomogeneity. Compared with the TPS, the absolute value of the transit dose at the beam center differed by -0.38% ± 2.1%. The simulation study indicated that the passing rate of the gamma index was significantly reduced, to less than 40%, when a wrong field was erroneously irradiated to patient in the treatment room. Conclusions: This feasibility study suggested that transit dosimetry based on the calculation with commercial TPS and EPID measurement with simple calibration can provide information about large errors for treatment beam delivery.

AB - Purpose: The aim of this study is to evaluate the ability of transit dosimetry using commercial treatment planning system (TPS) and an electronic portal imaging device (EPID) with simple calibration method to verify the beam delivery based on detection of large errors in treatment room. Methods and materials: Twenty four fields of intensity modulated radiotherapy (IMRT) plans were selected from four lung cancer patients and used in the irradiation of an anthropomorphic phantom. The proposed method was evaluated by comparing the calculated dose map from TPS and EPID measurement on the same plane using a gamma index method with a 3% dose and 3 mm distance-to-dose agreement tolerance limit. Results: In a simulation using a homogeneous plastic water phantom, performed to verify the effectiveness of the proposed method, the average passing rate of the transit dose based on gamma index was high enough, averaging 94.2% when there was no error during beam delivery. The passing rate of the transit dose for 24 IMRT fields was lower with the anthropomorphic phantom, averaging 86.8% ± 3.8%, a reduction partially due to the inaccuracy of TPS calculations for inhomogeneity. Compared with the TPS, the absolute value of the transit dose at the beam center differed by -0.38% ± 2.1%. The simulation study indicated that the passing rate of the gamma index was significantly reduced, to less than 40%, when a wrong field was erroneously irradiated to patient in the treatment room. Conclusions: This feasibility study suggested that transit dosimetry based on the calculation with commercial TPS and EPID measurement with simple calibration can provide information about large errors for treatment beam delivery.

KW - EPID

KW - Gamma index

KW - Intensity modulated radiation therapy

KW - Transit dose

UR - http://www.scopus.com/inward/record.url?scp=84924093272&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84924093272&partnerID=8YFLogxK

U2 - 10.1186/s13014-014-0273-8

DO - 10.1186/s13014-014-0273-8

M3 - Article

C2 - 25472838

AN - SCOPUS:84924093272

VL - 9

JO - Radiation Oncology

JF - Radiation Oncology

SN - 1748-717X

IS - 1

M1 - 273

ER -