New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study

Jung Su Kim, Soon Mu Kwon, Jung Min Kim, Sang Wook Yoon

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

A new organ-based tube current modulation (NOB-TCM) method was designed with the intent to decrease tube current by 30% over a prescribed 90° radial arc across the anterior aspect of the radiosensitive organ, without increasing tube current in the remaining radial arc. We compared a reference scan and five other dose-reducing methods with regard to effects on dose, practicality, and image quality to determine the most effective method for the reduction of the radiation dose to the eyes during CT examinations of the head. We compared the radiation doses to the eyes and physical image quality in different regions of interest for TCM and shielding scans. Three types of TCM scans were performed: longitudinal TCM, angular TCM, and NOB-TCM. A bismuth sheet and lead goggles were each applied for the shielding scan. Relative to the reference scan, the dose to the eye was reduced to 25.88% with NOB-TCM, 44.53% with lead goggles, and 36.91% with a bismuth shield. Relative to the reference scan, the mean signal-to-noise ratio (SNR) was decreased to 8.02% with NOB-TCM, 28.36% with lead goggles, and 32.95% with the bismuth shield. The SNR of the anterior region of interest was decreased to 11.89% with NOB-TCM and 87.89% with the bismuth shield. The average figure of merit was increased by 11.7% with longitudinal TCM and 13.39% with NOB-TCM, compared with the reference scan. NOB-TCM is a superior solution for head CT, including the orbital area, due to the reduction in radiation exposure without significant loss in image quality.

Original languageEnglish
Pages (from-to)601-608
Number of pages8
JournalRadiologia Medica
Volume122
Issue number8
DOIs
Publication statusPublished - 2017 Aug 1

Fingerprint

Eye Protective Devices
Bismuth
Head
Tomography
Signal-To-Noise Ratio
Radiation
Prefrontal Cortex
Lead

Keywords

  • Computed tomography
  • Dose reduction
  • Image quality
  • New organ-based tube current modulation
  • Radiation dose

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

@article{d54b0c08f5784d14ad238d290b243435,
title = "New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study",
abstract = "A new organ-based tube current modulation (NOB-TCM) method was designed with the intent to decrease tube current by 30{\%} over a prescribed 90° radial arc across the anterior aspect of the radiosensitive organ, without increasing tube current in the remaining radial arc. We compared a reference scan and five other dose-reducing methods with regard to effects on dose, practicality, and image quality to determine the most effective method for the reduction of the radiation dose to the eyes during CT examinations of the head. We compared the radiation doses to the eyes and physical image quality in different regions of interest for TCM and shielding scans. Three types of TCM scans were performed: longitudinal TCM, angular TCM, and NOB-TCM. A bismuth sheet and lead goggles were each applied for the shielding scan. Relative to the reference scan, the dose to the eye was reduced to 25.88{\%} with NOB-TCM, 44.53{\%} with lead goggles, and 36.91{\%} with a bismuth shield. Relative to the reference scan, the mean signal-to-noise ratio (SNR) was decreased to 8.02{\%} with NOB-TCM, 28.36{\%} with lead goggles, and 32.95{\%} with the bismuth shield. The SNR of the anterior region of interest was decreased to 11.89{\%} with NOB-TCM and 87.89{\%} with the bismuth shield. The average figure of merit was increased by 11.7{\%} with longitudinal TCM and 13.39{\%} with NOB-TCM, compared with the reference scan. NOB-TCM is a superior solution for head CT, including the orbital area, due to the reduction in radiation exposure without significant loss in image quality.",
keywords = "Computed tomography, Dose reduction, Image quality, New organ-based tube current modulation, Radiation dose",
author = "Kim, {Jung Su} and Kwon, {Soon Mu} and Kim, {Jung Min} and Yoon, {Sang Wook}",
year = "2017",
month = "8",
day = "1",
doi = "10.1007/s11547-017-0755-5",
language = "English",
volume = "122",
pages = "601--608",
journal = "Radiologia Medica",
issn = "0033-8362",
publisher = "Springer-Verlag Italia",
number = "8",

}

TY - JOUR

T1 - New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head

T2 - evaluation of image quality and radiation dose to the eyes in the phantom study

AU - Kim, Jung Su

AU - Kwon, Soon Mu

AU - Kim, Jung Min

AU - Yoon, Sang Wook

PY - 2017/8/1

Y1 - 2017/8/1

N2 - A new organ-based tube current modulation (NOB-TCM) method was designed with the intent to decrease tube current by 30% over a prescribed 90° radial arc across the anterior aspect of the radiosensitive organ, without increasing tube current in the remaining radial arc. We compared a reference scan and five other dose-reducing methods with regard to effects on dose, practicality, and image quality to determine the most effective method for the reduction of the radiation dose to the eyes during CT examinations of the head. We compared the radiation doses to the eyes and physical image quality in different regions of interest for TCM and shielding scans. Three types of TCM scans were performed: longitudinal TCM, angular TCM, and NOB-TCM. A bismuth sheet and lead goggles were each applied for the shielding scan. Relative to the reference scan, the dose to the eye was reduced to 25.88% with NOB-TCM, 44.53% with lead goggles, and 36.91% with a bismuth shield. Relative to the reference scan, the mean signal-to-noise ratio (SNR) was decreased to 8.02% with NOB-TCM, 28.36% with lead goggles, and 32.95% with the bismuth shield. The SNR of the anterior region of interest was decreased to 11.89% with NOB-TCM and 87.89% with the bismuth shield. The average figure of merit was increased by 11.7% with longitudinal TCM and 13.39% with NOB-TCM, compared with the reference scan. NOB-TCM is a superior solution for head CT, including the orbital area, due to the reduction in radiation exposure without significant loss in image quality.

AB - A new organ-based tube current modulation (NOB-TCM) method was designed with the intent to decrease tube current by 30% over a prescribed 90° radial arc across the anterior aspect of the radiosensitive organ, without increasing tube current in the remaining radial arc. We compared a reference scan and five other dose-reducing methods with regard to effects on dose, practicality, and image quality to determine the most effective method for the reduction of the radiation dose to the eyes during CT examinations of the head. We compared the radiation doses to the eyes and physical image quality in different regions of interest for TCM and shielding scans. Three types of TCM scans were performed: longitudinal TCM, angular TCM, and NOB-TCM. A bismuth sheet and lead goggles were each applied for the shielding scan. Relative to the reference scan, the dose to the eye was reduced to 25.88% with NOB-TCM, 44.53% with lead goggles, and 36.91% with a bismuth shield. Relative to the reference scan, the mean signal-to-noise ratio (SNR) was decreased to 8.02% with NOB-TCM, 28.36% with lead goggles, and 32.95% with the bismuth shield. The SNR of the anterior region of interest was decreased to 11.89% with NOB-TCM and 87.89% with the bismuth shield. The average figure of merit was increased by 11.7% with longitudinal TCM and 13.39% with NOB-TCM, compared with the reference scan. NOB-TCM is a superior solution for head CT, including the orbital area, due to the reduction in radiation exposure without significant loss in image quality.

KW - Computed tomography

KW - Dose reduction

KW - Image quality

KW - New organ-based tube current modulation

KW - Radiation dose

UR - http://www.scopus.com/inward/record.url?scp=85016064276&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85016064276&partnerID=8YFLogxK

U2 - 10.1007/s11547-017-0755-5

DO - 10.1007/s11547-017-0755-5

M3 - Article

C2 - 28341967

AN - SCOPUS:85016064276

VL - 122

SP - 601

EP - 608

JO - Radiologia Medica

JF - Radiologia Medica

SN - 0033-8362

IS - 8

ER -