Operation and performance analyses of 350 and 700 MHz low-/high-temperature superconductor nuclear magnetic resonance magnets: A march toward operating frequencies above 1 GHz

Seungyong Hahn, Juan Bascuán, Haigun Lee, Emanuel S. Bobrov, Wooseok Kim, Min Cheol Ahn, Yukikazu Iwasa

Research output: Contribution to journalArticle

33 Citations (Scopus)

Abstract

Since 2000, a three-phase program with a final goal to complete a 1 GHz high-resolution low-/high-temperature superconductor (LTS/HTS) nuclear magnetic resonance (NMR) magnet has been conducted at the Francis Bitter Magnet Laboratory, Massachusetts Institute of Technology (MIT). In a LTS/HTS magnet assembly, a HTS insert is placed in the cold bore of a LTS background magnet. To date, two LTS/HTS magnets have been designed, constructed, and tested: a 350 MHz (LH350) in phase 1 and a 700 MHz (LH700) in phase 2. The program's target has recently been upgraded from the original goal of 1 GHz to a new goal of 1.3 GHz. In this paper, we present extensive performance analyses of the two LTS/HTS NMR magnets. Spatial homogeneity and temporal stability of LH350 and LH700, examined with harmonic analysis, and four key issues that became evident in the operation of these two magnets are discussed: (1) field constant reduction, (2) "large" residual Z1 gradient and its temporal decay, (3) large one-periodic tesseral field gradients, and (4) screening-current-induced field in the HTS inserts.

Original languageEnglish
Article number024501
JournalJournal of Applied Physics
Volume105
Issue number2
DOIs
Publication statusPublished - 2009 Feb 9

Fingerprint

high temperature superconductors
magnets
nuclear magnetic resonance
inserts
gradients
harmonic analysis
homogeneity
screening
assembly
cavities
high resolution
decay

ASJC Scopus subject areas

  • Physics and Astronomy(all)

Cite this

Operation and performance analyses of 350 and 700 MHz low-/high-temperature superconductor nuclear magnetic resonance magnets : A march toward operating frequencies above 1 GHz. / Hahn, Seungyong; Bascuán, Juan; Lee, Haigun; Bobrov, Emanuel S.; Kim, Wooseok; Ahn, Min Cheol; Iwasa, Yukikazu.

In: Journal of Applied Physics, Vol. 105, No. 2, 024501, 09.02.2009.

Research output: Contribution to journalArticle

@article{6bd8791e63234abc992b2ae54da1cab4,
title = "Operation and performance analyses of 350 and 700 MHz low-/high-temperature superconductor nuclear magnetic resonance magnets: A march toward operating frequencies above 1 GHz",
abstract = "Since 2000, a three-phase program with a final goal to complete a 1 GHz high-resolution low-/high-temperature superconductor (LTS/HTS) nuclear magnetic resonance (NMR) magnet has been conducted at the Francis Bitter Magnet Laboratory, Massachusetts Institute of Technology (MIT). In a LTS/HTS magnet assembly, a HTS insert is placed in the cold bore of a LTS background magnet. To date, two LTS/HTS magnets have been designed, constructed, and tested: a 350 MHz (LH350) in phase 1 and a 700 MHz (LH700) in phase 2. The program's target has recently been upgraded from the original goal of 1 GHz to a new goal of 1.3 GHz. In this paper, we present extensive performance analyses of the two LTS/HTS NMR magnets. Spatial homogeneity and temporal stability of LH350 and LH700, examined with harmonic analysis, and four key issues that became evident in the operation of these two magnets are discussed: (1) field constant reduction, (2) {"}large{"} residual Z1 gradient and its temporal decay, (3) large one-periodic tesseral field gradients, and (4) screening-current-induced field in the HTS inserts.",
author = "Seungyong Hahn and Juan Bascu{\'a}n and Haigun Lee and Bobrov, {Emanuel S.} and Wooseok Kim and Ahn, {Min Cheol} and Yukikazu Iwasa",
year = "2009",
month = "2",
day = "9",
doi = "10.1063/1.3065538",
language = "English",
volume = "105",
journal = "Journal of Applied Physics",
issn = "0021-8979",
publisher = "American Institute of Physics Publising LLC",
number = "2",

}

TY - JOUR

T1 - Operation and performance analyses of 350 and 700 MHz low-/high-temperature superconductor nuclear magnetic resonance magnets

T2 - A march toward operating frequencies above 1 GHz

AU - Hahn, Seungyong

AU - Bascuán, Juan

AU - Lee, Haigun

AU - Bobrov, Emanuel S.

AU - Kim, Wooseok

AU - Ahn, Min Cheol

AU - Iwasa, Yukikazu

PY - 2009/2/9

Y1 - 2009/2/9

N2 - Since 2000, a three-phase program with a final goal to complete a 1 GHz high-resolution low-/high-temperature superconductor (LTS/HTS) nuclear magnetic resonance (NMR) magnet has been conducted at the Francis Bitter Magnet Laboratory, Massachusetts Institute of Technology (MIT). In a LTS/HTS magnet assembly, a HTS insert is placed in the cold bore of a LTS background magnet. To date, two LTS/HTS magnets have been designed, constructed, and tested: a 350 MHz (LH350) in phase 1 and a 700 MHz (LH700) in phase 2. The program's target has recently been upgraded from the original goal of 1 GHz to a new goal of 1.3 GHz. In this paper, we present extensive performance analyses of the two LTS/HTS NMR magnets. Spatial homogeneity and temporal stability of LH350 and LH700, examined with harmonic analysis, and four key issues that became evident in the operation of these two magnets are discussed: (1) field constant reduction, (2) "large" residual Z1 gradient and its temporal decay, (3) large one-periodic tesseral field gradients, and (4) screening-current-induced field in the HTS inserts.

AB - Since 2000, a three-phase program with a final goal to complete a 1 GHz high-resolution low-/high-temperature superconductor (LTS/HTS) nuclear magnetic resonance (NMR) magnet has been conducted at the Francis Bitter Magnet Laboratory, Massachusetts Institute of Technology (MIT). In a LTS/HTS magnet assembly, a HTS insert is placed in the cold bore of a LTS background magnet. To date, two LTS/HTS magnets have been designed, constructed, and tested: a 350 MHz (LH350) in phase 1 and a 700 MHz (LH700) in phase 2. The program's target has recently been upgraded from the original goal of 1 GHz to a new goal of 1.3 GHz. In this paper, we present extensive performance analyses of the two LTS/HTS NMR magnets. Spatial homogeneity and temporal stability of LH350 and LH700, examined with harmonic analysis, and four key issues that became evident in the operation of these two magnets are discussed: (1) field constant reduction, (2) "large" residual Z1 gradient and its temporal decay, (3) large one-periodic tesseral field gradients, and (4) screening-current-induced field in the HTS inserts.

UR - http://www.scopus.com/inward/record.url?scp=59349087549&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=59349087549&partnerID=8YFLogxK

U2 - 10.1063/1.3065538

DO - 10.1063/1.3065538

M3 - Article

AN - SCOPUS:59349087549

VL - 105

JO - Journal of Applied Physics

JF - Journal of Applied Physics

SN - 0021-8979

IS - 2

M1 - 024501

ER -