Performance evaluation of the Anysis-200 platelet function analyzer in cardiac patients

Seong Jun Park, Jung Yoon, Hong Seog Seo, Chae Seung Lim

Research output: Contribution to journalArticlepeer-review

3 Citations (Scopus)

Abstract

BACKGROUND: Platelet function analysis is crucial in assessing the hemostatic status to evaluate congenital and acquired platelet function defects. The Anysis-200 analyzer is a new automated lab-on-a-chip-based platelet function analyzer. OBJECTIVE: We aimed to evaluate a new platelet function analyzing system, the Anysis-200 in comparison to the Platelet Function Analyzer (PFA)-200 in cardiac patients. METHODS: Citrated blood was collected from 174 patients who visited the Department of Cardiology. The Anysis-200 consists of two kits, the microchips with collagen and epinephrine-coated membrane (C/EPI) or adenosine diphosphate-coated membrane (C/ADP). Platelet clogging in the Anysis-200 is measured by the blood migration distance obtained by a camera, which is compatible with the closure time in the PFA-200. We performed Anysis-200 and PFA-200 analyzers simultaneously and compared the results. RESULTS: The sensitivity and specificity of the Anysis-200 C/EPI kit in comparison to the PFA-200 C/EPI kit were 63.41% and 91.43%, respectively. Regarding the C/ADP kit, the sensitivity and specificity of the Anysis-200 were 58.97% and 74.29%, respectively. The agreement rate between the Anysis-200 and PFA-200 for C/EPI was 83.35% and 70.14% for C/ADP. CONCLUSIONS: The Anysis-200, which applies a novel method to detect platelet clogging, has shown moderate to fair agreement with the PFA-200. This test is potentially useful for screening cardiac patients with an abnormal platelet function.

Original languageEnglish
Pages (from-to)17-24
Number of pages8
JournalClinical Hemorheology and Microcirculation
Volume80
Issue number1
DOIs
Publication statusPublished - 2022

Keywords

  • Anysis-200 analyzer
  • PFA-200 analyzer
  • Platelet function

ASJC Scopus subject areas

  • Physiology
  • Hematology
  • Cardiology and Cardiovascular Medicine
  • Physiology (medical)

Fingerprint

Dive into the research topics of 'Performance evaluation of the Anysis-200 platelet function analyzer in cardiac patients'. Together they form a unique fingerprint.

Cite this