Abstract

Case: We present two cases of anterior cruciate ligament reconstruction with use of the flexible reamer system. In both patients, the starting reamers were broken during femoral reaming. The location of the breaks was approximately 60 mm from the tip of the reamer, which is the transitional zone between the proximal and distal portions of the reamer. Conclusion: We believe that several conditions could have resulted in abnormal stress concentration on the flexible reamers, including engagement of the flexible reamer with a physeal scar or the posterior cortex, an inappropriate flexion angle of the knee, and acute bending of the flexible pin.

Original languageEnglish
Article numbere95
JournalJBJS Case Connector
Volume4
Issue number4
DOIs
Publication statusPublished - 2014 Jan 1

Fingerprint

Anterior Cruciate Ligament Reconstruction
Thigh
Cicatrix
Knee

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Cite this

@article{0cddb791f34e411eaaa08b9dee607565,
title = "Pitfalls in single-bundle anterior cruciate ligament reconstruction with the flexible reamer system: A report of two cases and review of the literature",
abstract = "Case: We present two cases of anterior cruciate ligament reconstruction with use of the flexible reamer system. In both patients, the starting reamers were broken during femoral reaming. The location of the breaks was approximately 60 mm from the tip of the reamer, which is the transitional zone between the proximal and distal portions of the reamer. Conclusion: We believe that several conditions could have resulted in abnormal stress concentration on the flexible reamers, including engagement of the flexible reamer with a physeal scar or the posterior cortex, an inappropriate flexion angle of the knee, and acute bending of the flexible pin.",
author = "Lee, {Seung Yup} and Kim, {Jae Gyoon} and Ki-Mo Jang and Lim, {Hong Chul}",
year = "2014",
month = "1",
day = "1",
doi = "10.2106/JBJS.CC.N.00045",
language = "English",
volume = "4",
journal = "JBJS Case Connector",
issn = "2160-3251",
publisher = "Lippincott Williams and Wilkins",
number = "4",

}

TY - JOUR

T1 - Pitfalls in single-bundle anterior cruciate ligament reconstruction with the flexible reamer system

T2 - A report of two cases and review of the literature

AU - Lee, Seung Yup

AU - Kim, Jae Gyoon

AU - Jang, Ki-Mo

AU - Lim, Hong Chul

PY - 2014/1/1

Y1 - 2014/1/1

N2 - Case: We present two cases of anterior cruciate ligament reconstruction with use of the flexible reamer system. In both patients, the starting reamers were broken during femoral reaming. The location of the breaks was approximately 60 mm from the tip of the reamer, which is the transitional zone between the proximal and distal portions of the reamer. Conclusion: We believe that several conditions could have resulted in abnormal stress concentration on the flexible reamers, including engagement of the flexible reamer with a physeal scar or the posterior cortex, an inappropriate flexion angle of the knee, and acute bending of the flexible pin.

AB - Case: We present two cases of anterior cruciate ligament reconstruction with use of the flexible reamer system. In both patients, the starting reamers were broken during femoral reaming. The location of the breaks was approximately 60 mm from the tip of the reamer, which is the transitional zone between the proximal and distal portions of the reamer. Conclusion: We believe that several conditions could have resulted in abnormal stress concentration on the flexible reamers, including engagement of the flexible reamer with a physeal scar or the posterior cortex, an inappropriate flexion angle of the knee, and acute bending of the flexible pin.

UR - http://www.scopus.com/inward/record.url?scp=85027542545&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85027542545&partnerID=8YFLogxK

U2 - 10.2106/JBJS.CC.N.00045

DO - 10.2106/JBJS.CC.N.00045

M3 - Article

VL - 4

JO - JBJS Case Connector

JF - JBJS Case Connector

SN - 2160-3251

IS - 4

M1 - e95

ER -