In proton therapy, patient quality assurance (QA) requires measuring the beam range, spread-out Bragg peak (SOBP), and output factor. If these values can be predicted by using sampling measurements or previous QA data to find the correlation between beam setup parameters and measured data, efforts expended on patient QA can be reduced. Using sampling data, we predicted the range, SOBP, and output factor of the proton beam. To obtain sampling data, we measured the range, SOBP, and output factor for 14 data points at each of 24-beam range options, from 4-28 cm. Prediction conformity was evaluated by the difference between predicted and measured patient QA data. Results indicated that for 60% of patients, the values could be predicted within 3% of dose uncertainty.

Original languageEnglish
Pages (from-to)145-152
Number of pages8
JournalMedical Dosimetry
Volume36
Issue number2
DOIs
Publication statusPublished - 2011 Jan 1
Externally publishedYes

Fingerprint

Proton Therapy
Uncertainty
Protons
Data Accuracy

Keywords

  • Dose
  • Output
  • Prediction
  • Proton
  • Range
  • SOBP

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Prediction of output factor, range, and spread-out bragg peak for proton therapy. / Kim, Dong Wook; Lim, Young Kyung; Ahn, Sung Hwan; Shin, Jungwook; Shin, Dongho; Yoon, Myonggeun; Lee, Se Byeong; Kim, Dae Yong; Park, Sung Yong.

In: Medical Dosimetry, Vol. 36, No. 2, 01.01.2011, p. 145-152.

Research output: Contribution to journalArticle

Kim, DW, Lim, YK, Ahn, SH, Shin, J, Shin, D, Yoon, M, Lee, SB, Kim, DY & Park, SY 2011, 'Prediction of output factor, range, and spread-out bragg peak for proton therapy', Medical Dosimetry, vol. 36, no. 2, pp. 145-152. https://doi.org/10.1016/j.meddos.2010.02.006
Kim, Dong Wook ; Lim, Young Kyung ; Ahn, Sung Hwan ; Shin, Jungwook ; Shin, Dongho ; Yoon, Myonggeun ; Lee, Se Byeong ; Kim, Dae Yong ; Park, Sung Yong. / Prediction of output factor, range, and spread-out bragg peak for proton therapy. In: Medical Dosimetry. 2011 ; Vol. 36, No. 2. pp. 145-152.
@article{f439917609a0412ea54612141ef07254,
title = "Prediction of output factor, range, and spread-out bragg peak for proton therapy",
abstract = "In proton therapy, patient quality assurance (QA) requires measuring the beam range, spread-out Bragg peak (SOBP), and output factor. If these values can be predicted by using sampling measurements or previous QA data to find the correlation between beam setup parameters and measured data, efforts expended on patient QA can be reduced. Using sampling data, we predicted the range, SOBP, and output factor of the proton beam. To obtain sampling data, we measured the range, SOBP, and output factor for 14 data points at each of 24-beam range options, from 4-28 cm. Prediction conformity was evaluated by the difference between predicted and measured patient QA data. Results indicated that for 60{\%} of patients, the values could be predicted within 3{\%} of dose uncertainty.",
keywords = "Dose, Output, Prediction, Proton, Range, SOBP",
author = "Kim, {Dong Wook} and Lim, {Young Kyung} and Ahn, {Sung Hwan} and Jungwook Shin and Dongho Shin and Myonggeun Yoon and Lee, {Se Byeong} and Kim, {Dae Yong} and Park, {Sung Yong}",
year = "2011",
month = "1",
day = "1",
doi = "10.1016/j.meddos.2010.02.006",
language = "English",
volume = "36",
pages = "145--152",
journal = "Medical Dosimetry",
issn = "0958-3947",
publisher = "Elsevier Inc.",
number = "2",

}

TY - JOUR

T1 - Prediction of output factor, range, and spread-out bragg peak for proton therapy

AU - Kim, Dong Wook

AU - Lim, Young Kyung

AU - Ahn, Sung Hwan

AU - Shin, Jungwook

AU - Shin, Dongho

AU - Yoon, Myonggeun

AU - Lee, Se Byeong

AU - Kim, Dae Yong

AU - Park, Sung Yong

PY - 2011/1/1

Y1 - 2011/1/1

N2 - In proton therapy, patient quality assurance (QA) requires measuring the beam range, spread-out Bragg peak (SOBP), and output factor. If these values can be predicted by using sampling measurements or previous QA data to find the correlation between beam setup parameters and measured data, efforts expended on patient QA can be reduced. Using sampling data, we predicted the range, SOBP, and output factor of the proton beam. To obtain sampling data, we measured the range, SOBP, and output factor for 14 data points at each of 24-beam range options, from 4-28 cm. Prediction conformity was evaluated by the difference between predicted and measured patient QA data. Results indicated that for 60% of patients, the values could be predicted within 3% of dose uncertainty.

AB - In proton therapy, patient quality assurance (QA) requires measuring the beam range, spread-out Bragg peak (SOBP), and output factor. If these values can be predicted by using sampling measurements or previous QA data to find the correlation between beam setup parameters and measured data, efforts expended on patient QA can be reduced. Using sampling data, we predicted the range, SOBP, and output factor of the proton beam. To obtain sampling data, we measured the range, SOBP, and output factor for 14 data points at each of 24-beam range options, from 4-28 cm. Prediction conformity was evaluated by the difference between predicted and measured patient QA data. Results indicated that for 60% of patients, the values could be predicted within 3% of dose uncertainty.

KW - Dose

KW - Output

KW - Prediction

KW - Proton

KW - Range

KW - SOBP

UR - http://www.scopus.com/inward/record.url?scp=79955467759&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79955467759&partnerID=8YFLogxK

U2 - 10.1016/j.meddos.2010.02.006

DO - 10.1016/j.meddos.2010.02.006

M3 - Article

C2 - 20599372

AN - SCOPUS:79955467759

VL - 36

SP - 145

EP - 152

JO - Medical Dosimetry

JF - Medical Dosimetry

SN - 0958-3947

IS - 2

ER -