Stem diameter and micromotion of press fit radial head prosthesis: A biomechanical study

Jungyu Moon, Lawrence J. Berglund, Zachary Domire, Kai Nan An, Shawn W. O'Driscoll

Research output: Contribution to journalArticle

29 Citations (Scopus)

Abstract

Background: Aseptic loosening of the stem, ranging from mild periprosthetic lucency to symptomatic loosening leading to implant removal, has been reported in press fit radial head prostheses. Hypothesis: The purpose of this study was to determine the effect of the stem diameter and insertion force on initial stability with a press fit radial head prosthesis designed for bone ingrowth. Materials and methods: Cadaveric radii were implanted with radial head prostheses of increasing stem diameter. The insertion forces for each rasp and stem were measured. Micromotion of the stem at the isthmus of the canal and stem tip were measured under circumstances simulating eccentric loads. Results: Insertion forces for all submaximum-sized rasps were similar. However, the insertion force for the maximum-sized rasp was approximately twice as large, and the insertion force for the oversized rasp was twice as large again, potentially indicating that the insertion force may be useful as a guide for determining appropriate stem size. Micromotions of the maximum diameter stem (isthmus, 41 μm; tip, 64 μm) were near the threshold for bone ingrowth, whereas the micromotions of the submaximum stem (isthmus, 253 μm; tip, 394 μm) were above this threshold. Discussion: The maximum diameter stem achieved greater stability (minimum micromotion) compared with the submaximum diameter stem. Conclusion: The best fixation strength in the press fit radial head prosthesis was achieved by maximum sizing in the neck canal. Level of evidence: Basic science study.

Original languageEnglish
Pages (from-to)785-790
Number of pages6
JournalJournal of Shoulder and Elbow Surgery
Volume18
Issue number5
DOIs
Publication statusPublished - 2009 Sep 1
Externally publishedYes

Fingerprint

Prostheses and Implants
Bone and Bones
Neck

Keywords

  • initial stability
  • Radial head prosthesis
  • stem diameter

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine
  • Surgery

Cite this

Stem diameter and micromotion of press fit radial head prosthesis : A biomechanical study. / Moon, Jungyu; Berglund, Lawrence J.; Domire, Zachary; An, Kai Nan; O'Driscoll, Shawn W.

In: Journal of Shoulder and Elbow Surgery, Vol. 18, No. 5, 01.09.2009, p. 785-790.

Research output: Contribution to journalArticle

Moon, Jungyu ; Berglund, Lawrence J. ; Domire, Zachary ; An, Kai Nan ; O'Driscoll, Shawn W. / Stem diameter and micromotion of press fit radial head prosthesis : A biomechanical study. In: Journal of Shoulder and Elbow Surgery. 2009 ; Vol. 18, No. 5. pp. 785-790.
@article{d1c4f6227d104087bf2f384a47f07180,
title = "Stem diameter and micromotion of press fit radial head prosthesis: A biomechanical study",
abstract = "Background: Aseptic loosening of the stem, ranging from mild periprosthetic lucency to symptomatic loosening leading to implant removal, has been reported in press fit radial head prostheses. Hypothesis: The purpose of this study was to determine the effect of the stem diameter and insertion force on initial stability with a press fit radial head prosthesis designed for bone ingrowth. Materials and methods: Cadaveric radii were implanted with radial head prostheses of increasing stem diameter. The insertion forces for each rasp and stem were measured. Micromotion of the stem at the isthmus of the canal and stem tip were measured under circumstances simulating eccentric loads. Results: Insertion forces for all submaximum-sized rasps were similar. However, the insertion force for the maximum-sized rasp was approximately twice as large, and the insertion force for the oversized rasp was twice as large again, potentially indicating that the insertion force may be useful as a guide for determining appropriate stem size. Micromotions of the maximum diameter stem (isthmus, 41 μm; tip, 64 μm) were near the threshold for bone ingrowth, whereas the micromotions of the submaximum stem (isthmus, 253 μm; tip, 394 μm) were above this threshold. Discussion: The maximum diameter stem achieved greater stability (minimum micromotion) compared with the submaximum diameter stem. Conclusion: The best fixation strength in the press fit radial head prosthesis was achieved by maximum sizing in the neck canal. Level of evidence: Basic science study.",
keywords = "initial stability, Radial head prosthesis, stem diameter",
author = "Jungyu Moon and Berglund, {Lawrence J.} and Zachary Domire and An, {Kai Nan} and O'Driscoll, {Shawn W.}",
year = "2009",
month = "9",
day = "1",
doi = "10.1016/j.jse.2009.02.014",
language = "English",
volume = "18",
pages = "785--790",
journal = "Journal of Shoulder and Elbow Surgery",
issn = "1058-2746",
publisher = "Mosby Inc.",
number = "5",

}

TY - JOUR

T1 - Stem diameter and micromotion of press fit radial head prosthesis

T2 - A biomechanical study

AU - Moon, Jungyu

AU - Berglund, Lawrence J.

AU - Domire, Zachary

AU - An, Kai Nan

AU - O'Driscoll, Shawn W.

PY - 2009/9/1

Y1 - 2009/9/1

N2 - Background: Aseptic loosening of the stem, ranging from mild periprosthetic lucency to symptomatic loosening leading to implant removal, has been reported in press fit radial head prostheses. Hypothesis: The purpose of this study was to determine the effect of the stem diameter and insertion force on initial stability with a press fit radial head prosthesis designed for bone ingrowth. Materials and methods: Cadaveric radii were implanted with radial head prostheses of increasing stem diameter. The insertion forces for each rasp and stem were measured. Micromotion of the stem at the isthmus of the canal and stem tip were measured under circumstances simulating eccentric loads. Results: Insertion forces for all submaximum-sized rasps were similar. However, the insertion force for the maximum-sized rasp was approximately twice as large, and the insertion force for the oversized rasp was twice as large again, potentially indicating that the insertion force may be useful as a guide for determining appropriate stem size. Micromotions of the maximum diameter stem (isthmus, 41 μm; tip, 64 μm) were near the threshold for bone ingrowth, whereas the micromotions of the submaximum stem (isthmus, 253 μm; tip, 394 μm) were above this threshold. Discussion: The maximum diameter stem achieved greater stability (minimum micromotion) compared with the submaximum diameter stem. Conclusion: The best fixation strength in the press fit radial head prosthesis was achieved by maximum sizing in the neck canal. Level of evidence: Basic science study.

AB - Background: Aseptic loosening of the stem, ranging from mild periprosthetic lucency to symptomatic loosening leading to implant removal, has been reported in press fit radial head prostheses. Hypothesis: The purpose of this study was to determine the effect of the stem diameter and insertion force on initial stability with a press fit radial head prosthesis designed for bone ingrowth. Materials and methods: Cadaveric radii were implanted with radial head prostheses of increasing stem diameter. The insertion forces for each rasp and stem were measured. Micromotion of the stem at the isthmus of the canal and stem tip were measured under circumstances simulating eccentric loads. Results: Insertion forces for all submaximum-sized rasps were similar. However, the insertion force for the maximum-sized rasp was approximately twice as large, and the insertion force for the oversized rasp was twice as large again, potentially indicating that the insertion force may be useful as a guide for determining appropriate stem size. Micromotions of the maximum diameter stem (isthmus, 41 μm; tip, 64 μm) were near the threshold for bone ingrowth, whereas the micromotions of the submaximum stem (isthmus, 253 μm; tip, 394 μm) were above this threshold. Discussion: The maximum diameter stem achieved greater stability (minimum micromotion) compared with the submaximum diameter stem. Conclusion: The best fixation strength in the press fit radial head prosthesis was achieved by maximum sizing in the neck canal. Level of evidence: Basic science study.

KW - initial stability

KW - Radial head prosthesis

KW - stem diameter

UR - http://www.scopus.com/inward/record.url?scp=68749120633&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=68749120633&partnerID=8YFLogxK

U2 - 10.1016/j.jse.2009.02.014

DO - 10.1016/j.jse.2009.02.014

M3 - Article

C2 - 19427236

AN - SCOPUS:68749120633

VL - 18

SP - 785

EP - 790

JO - Journal of Shoulder and Elbow Surgery

JF - Journal of Shoulder and Elbow Surgery

SN - 1058-2746

IS - 5

ER -