Technical difficulties of removal of locking screw after locking compression plating

Ji Hoon Bae, Jong Keon Oh, Chang Wug Oh, Chang Ryong Hur

    Research output: Contribution to journalArticlepeer-review

    48 Citations (Scopus)

    Abstract

    Introduction: As there are few reports on the difficulties of removing the locking compression plate (LCP), we prospectively investigated the incidence and difficulties in 58 patients in whom various types of LCPs were taken. Methods: From January 2004 to December 2007, we have removed 159 5.0-self tapping locking screws and 279 3.5- self tapping locking screws. All of the operations were performed by experienced trauma surgeons. All of the screws were inserted with the use of torque limiting attachment according to the manufacturers recommendation. During the same period of time we have removed 198 AO-3.5 cortical and 4.0 cancellous screws from various sites. Results: All of 159 5.0-self tapping locking screws were removed without difficulties. A total of 24 out of 279 3.5- self tapping locking screws were removed with many difficulties due to the stripping of the hexagonal recess. The use of conical extraction screw which was developed especially for the removal of stripped locking screws was successful in only six screws. We have removed plates by cutting the plate with metal cutting saw. We describe useful technical trick to remove the plate when there is only one screw left stripped. Compared to the locking screws, only one of 198 3.5-cortical screws was stripped. Conclusion: Care should be taken at the time of removal of the locking compression plate, especially for the 3.5-locking screws.

    Original languageEnglish
    Pages (from-to)91-95
    Number of pages5
    JournalArchives of Orthopaedic and Trauma Surgery
    Volume129
    Issue number1
    DOIs
    Publication statusPublished - 2009

    Keywords

    • Locking compression plate
    • Locking screw
    • Removal
    • Stripping
    • Technique

    ASJC Scopus subject areas

    • Surgery
    • Orthopedics and Sports Medicine

    Fingerprint

    Dive into the research topics of 'Technical difficulties of removal of locking screw after locking compression plating'. Together they form a unique fingerprint.

    Cite this