T1 - Tenelectrodes
T2 - A new stimulator for inching technique in the diagnosis of carpal tunnel syndrome
AU - Kang, Yoon Kyoo
AU - Kim, Dong Hwee
AU - Lee, Seung Hwa
AU - Hwang, Miriam
AU - Han, Myung Soo
N1 - Funding Information:
The authors would like to thank the Air Force Office of Scientific Research for the support of this study.
PY - 2003/6/30
Y1 - 2003/6/30
N2 - This study was designed evaluate the usefulness of a new multielectrode stimulator, TenElectrodes, in the diagnosis and localization of the compression site in the wrists of carpal tunnel syndrome (CTS) patients. Antidromic inching technique (IT) of the median nerve at the wrist was performed with the TenElectrodes, on 46 controls and 21 CTS patients. In controls, mean conduction delay per centimeter (CD/cm) was 0.21 milliseconds (ms), and maximal CD/cm was 0.27 ms in the segment 3 to 4 centimeters distal to the distal wrist crease. The abnormal cut-off value, calculated as the maximal CD/cm + 2SD, was 0.45 ms. In the CTS group, the maximal CD/cm was 0.56 ms in the segment 2 to 3 centimeters distal to the distal wrist crease, and the CD/cm values in all segments between the distal wrist crease and 4 cm distal to the distal wrist crease were greater than 0.45 ms. Antidromic IT using TenElectrodes may be an easy, fast and accurate method as the electrodes of the stimulator are aligned at 1-cm intervals and are adjustable to the wrist contour by springs.
AB - This study was designed evaluate the usefulness of a new multielectrode stimulator, TenElectrodes, in the diagnosis and localization of the compression site in the wrists of carpal tunnel syndrome (CTS) patients. Antidromic inching technique (IT) of the median nerve at the wrist was performed with the TenElectrodes, on 46 controls and 21 CTS patients. In controls, mean conduction delay per centimeter (CD/cm) was 0.21 milliseconds (ms), and maximal CD/cm was 0.27 ms in the segment 3 to 4 centimeters distal to the distal wrist crease. The abnormal cut-off value, calculated as the maximal CD/cm + 2SD, was 0.45 ms. In the CTS group, the maximal CD/cm was 0.56 ms in the segment 2 to 3 centimeters distal to the distal wrist crease, and the CD/cm values in all segments between the distal wrist crease and 4 cm distal to the distal wrist crease were greater than 0.45 ms. Antidromic IT using TenElectrodes may be an easy, fast and accurate method as the electrodes of the stimulator are aligned at 1-cm intervals and are adjustable to the wrist contour by springs.
KW - Carpal tunnel syndrome
KW - Inching technique
KW - Median nerve
KW - Multielectrode stimulator
KW - Nerve conduction study