Three-dimensional radiochromic film dosimetry of proton clinical beams using a gafchromic EBT2 film array

Jinsung Kim, Myonggeun Yoon, Seonkyu Kim, Dongho Shin, Se Byeong Lee, Young Kyung Lim, Dong Wook Kim, Sung Yong Park

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

In this work, three-dimensional (3D) film-based proton beam measurements were used for the first time to verify the patient-specific radiation dose distribution, beam range and compensator shape. Three passively scattered proton beams and one uniform scanning proton beam were directed onto an acrylic phantom with inserted Gafchromic EBT films. The average gamma index for a comparison of the dose distributions was less than one for 97.2 % of all pixels from the passively scattered proton beams and 98.1 % of all pixels for the uniform scanning proton beams, with a 3 % dose and a 3 mm distance-to-dose agreement tolerance limit. The results also showed that the average percentage of points within the acceptance criteria for proton beam ranges was 94.6 % for the passively scattered proton beams. Both the dose distribution and the proton beam range determined by the 3D EBT film measurement agreed well with the planning system values.

Original languageEnglish
Article numberncs010
Pages (from-to)272-277
Number of pages6
JournalRadiation Protection Dosimetry
Volume151
Issue number2
DOIs
Publication statusPublished - 2012 Aug 1

Fingerprint

Film Dosimetry
proton beams
dosimeters
Protons
protons
dosage
pixels
scanning
compensators
acceptability
planning
Radiation

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology
  • Radiation
  • Public Health, Environmental and Occupational Health

Cite this

Three-dimensional radiochromic film dosimetry of proton clinical beams using a gafchromic EBT2 film array. / Kim, Jinsung; Yoon, Myonggeun; Kim, Seonkyu; Shin, Dongho; Lee, Se Byeong; Lim, Young Kyung; Kim, Dong Wook; Park, Sung Yong.

In: Radiation Protection Dosimetry, Vol. 151, No. 2, ncs010, 01.08.2012, p. 272-277.

Research output: Contribution to journalArticle

Kim, Jinsung ; Yoon, Myonggeun ; Kim, Seonkyu ; Shin, Dongho ; Lee, Se Byeong ; Lim, Young Kyung ; Kim, Dong Wook ; Park, Sung Yong. / Three-dimensional radiochromic film dosimetry of proton clinical beams using a gafchromic EBT2 film array. In: Radiation Protection Dosimetry. 2012 ; Vol. 151, No. 2. pp. 272-277.
@article{f217f8fc16b24adab967f9bf9252c861,
title = "Three-dimensional radiochromic film dosimetry of proton clinical beams using a gafchromic EBT2 film array",
abstract = "In this work, three-dimensional (3D) film-based proton beam measurements were used for the first time to verify the patient-specific radiation dose distribution, beam range and compensator shape. Three passively scattered proton beams and one uniform scanning proton beam were directed onto an acrylic phantom with inserted Gafchromic EBT films. The average gamma index for a comparison of the dose distributions was less than one for 97.2 {\%} of all pixels from the passively scattered proton beams and 98.1 {\%} of all pixels for the uniform scanning proton beams, with a 3 {\%} dose and a 3 mm distance-to-dose agreement tolerance limit. The results also showed that the average percentage of points within the acceptance criteria for proton beam ranges was 94.6 {\%} for the passively scattered proton beams. Both the dose distribution and the proton beam range determined by the 3D EBT film measurement agreed well with the planning system values.",
author = "Jinsung Kim and Myonggeun Yoon and Seonkyu Kim and Dongho Shin and Lee, {Se Byeong} and Lim, {Young Kyung} and Kim, {Dong Wook} and Park, {Sung Yong}",
year = "2012",
month = "8",
day = "1",
doi = "10.1093/rpd/ncs010",
language = "English",
volume = "151",
pages = "272--277",
journal = "Radiation Protection Dosimetry",
issn = "0144-8420",
publisher = "Oxford University Press",
number = "2",

}

TY - JOUR

T1 - Three-dimensional radiochromic film dosimetry of proton clinical beams using a gafchromic EBT2 film array

AU - Kim, Jinsung

AU - Yoon, Myonggeun

AU - Kim, Seonkyu

AU - Shin, Dongho

AU - Lee, Se Byeong

AU - Lim, Young Kyung

AU - Kim, Dong Wook

AU - Park, Sung Yong

PY - 2012/8/1

Y1 - 2012/8/1

N2 - In this work, three-dimensional (3D) film-based proton beam measurements were used for the first time to verify the patient-specific radiation dose distribution, beam range and compensator shape. Three passively scattered proton beams and one uniform scanning proton beam were directed onto an acrylic phantom with inserted Gafchromic EBT films. The average gamma index for a comparison of the dose distributions was less than one for 97.2 % of all pixels from the passively scattered proton beams and 98.1 % of all pixels for the uniform scanning proton beams, with a 3 % dose and a 3 mm distance-to-dose agreement tolerance limit. The results also showed that the average percentage of points within the acceptance criteria for proton beam ranges was 94.6 % for the passively scattered proton beams. Both the dose distribution and the proton beam range determined by the 3D EBT film measurement agreed well with the planning system values.

AB - In this work, three-dimensional (3D) film-based proton beam measurements were used for the first time to verify the patient-specific radiation dose distribution, beam range and compensator shape. Three passively scattered proton beams and one uniform scanning proton beam were directed onto an acrylic phantom with inserted Gafchromic EBT films. The average gamma index for a comparison of the dose distributions was less than one for 97.2 % of all pixels from the passively scattered proton beams and 98.1 % of all pixels for the uniform scanning proton beams, with a 3 % dose and a 3 mm distance-to-dose agreement tolerance limit. The results also showed that the average percentage of points within the acceptance criteria for proton beam ranges was 94.6 % for the passively scattered proton beams. Both the dose distribution and the proton beam range determined by the 3D EBT film measurement agreed well with the planning system values.

UR - http://www.scopus.com/inward/record.url?scp=84865554779&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84865554779&partnerID=8YFLogxK

U2 - 10.1093/rpd/ncs010

DO - 10.1093/rpd/ncs010

M3 - Article

C2 - 22355172

AN - SCOPUS:84865554779

VL - 151

SP - 272

EP - 277

JO - Radiation Protection Dosimetry

JF - Radiation Protection Dosimetry

SN - 0144-8420

IS - 2

M1 - ncs010

ER -